14
6.
Route 3/8" I.D. vinyl drain tubing to atmospherically
vented drain.
7.
Plug solenoid valve power cord (not shown) into
receptacle in control box "H."
8.
Plug still adapter kit cable into right side of dem-
ineralizer and receptacle in control box "H."
9.
Go to the Still Output Connections section on page
16 to complete the setup of your still.
In-House Treated Boiler Feed
See Figure 9. (Must have optional double solenoid valves:
Catalog No. 440236 (240V), RY798X2A (120V))
1.
Install customer supplied shutoff valves onto your
tap water source and your in-house treated water
source. Thread the solenoid valves onto your cus-
tomer supplied shutoffs, with the sides marked “in"
toward the shutoffs.
2.
Install the supplied 5/16" fitting to the “out” side of
the solenoid valve on your tap water source.
Connect the 5/16" I.D. tubing (from the condenser)
to this fitting.
F
EEDWATER
C
ONNECTIONS
Figure 9 In-house Treated Feed