6
Specifications
Dimensions and Clearance Requirements.
EASYpure LF dimensions - 12 7/8" W X 19" D X 17 1/8" H (32.7 cm X 48.3 cm X 43.5 cm).
Clearances:
Sides - 4" (10.2 cm) minimum to allow air flow.
Above - 12" (30.5 cm) minimum for removal of the top cover.
Cartridge replacement requires that you be able to access the back of the unit and open the cartridge
access door (total depth, unit + open door, = 34") (86.4 cm).
Feedwater Requirements
The EASYpure LF requires water pretreated by either distillation, deionization or reverse osmosis meeting the
following criteria.
TOC - Less than 1.0 ppm.
Turbidity - 1.0 N.T.U. maximum
Temperature - 40°F - 120°F (4.4°C - 48.8°C)
Pressure: Gravity - 100 PSIG
Resistivity (Minimum):
Distilled - 300,000 Ohms-cm
Deionized - 1.0 Megohms-cm
Reverse osmosis - 100,000 Ohms-cm
Product Water
Quality
Resistivity: ASTM Type I
TOC: Less than 15.0 ppb
Flow Rate: 1.5 LPM at 30 PSIG inlet, feedwater pressure at 60Hz and with a new final filter.
Electrical Requirements
The EASYpure LF is equipped with a power cord to be plugged into an electrical outlet of the appropriate volt-
age.
Model D7383 - 100 VAC +5%, -10%, 47-63 Hz.
Model D7381 - 120 VAX +5%, -10%, 47-63 Hz.
Model D7382-33 - 230 VAC +10%, -10%, 47-63 Hz.
Model D7382 - 240 VAC +10%, -10%, 47-63 Hz.
WARNING
Use a properly grounded electrical outlet of correct voltage and current handling capacity.
Environmental Conditions
Operating: 4°C - 49°C; 20% to 80% relative humidity, non-condensing. Installation Category II
(over-voltage) in accordance with IEC 664. Pollution Degree 2 in accordance with IEC 664.
Altitude limit: 5,000 meters.
Storage: -25°C to 65°C; 20% to 80% relative humidity.