18
3.
Depressurize system by opening the draw-off
valve.
4.
Open the cartridge access door in the rear of the unit
by sliding the latch down and pulling the door toward
you. The door will swing down.
5.
Remove exhausted cartridge in the right-hand position
by pulling it first up and then out.
6.
Remove a D50245 sanitization cartridge from its pack-
aging. Press the upper end cap of the D50245 saniti-
zation cartridge into the upper right position until it bot-
toms out.
7.
Lower the cartridge and insert the lower end cap into
the lower socket until it is firmly seated.
8.
Repeat steps 2 - 6 with the two empty cartridges sup-
plied with your EASYpure LF, placing them in the cen-
ter and left-hand positions.
9.
Close the cartridge access door. Remove the 0.2 mi-
cron filter and bell assembly.
10. Sanitize, install new cartridges and rinse according to
the instructions for Cartridge Installation and Rinse Up
in the Initial Operation section.
Fuse Replacement
1.
Turn the system off.
2.
Disconnect incoming water line at quick disconnect.
Depressurize system by opening outlet valve.
3.
Disconnect the EASYpure LF from the power supply.
4.
Remove the power cord from the power entry module,
and remove the pump interlock plug from the rear of
the unit.
5.
Remove the screws securing the EASYpure LF cover.
6.
Remove the cover by lifting it straight up.
7.
Pull out the fuse drawer located in the power entry
module.
WARNING
Depressurize system prior to opening cartridge
access door.
NOTE
If the EASYpure LF is wall mounted, remove
the power cord and rotate the EASYpure LF
until the cartridge access door faces forward
and the EASYpure LF locks into place.
NOTE
The two flanges on the end cap should be able
to slide down on each side of the keyway wall.
WARNING
Depressurize system prior to opening cartridge
access door.
M
AINTENANCE AND
S
ERVICING