8
Unpackaging and Installation
1.
Remove the unit from its shipping container. Ensure
that the tubing, adapter and power cord are removed
from the packaging materials before discarding.
Description
Catalog Number
EASYpure High Purity/Low TOC Cartridge
D50229
Pretreatment Cartridge DI Feed
D50230
Pretreatment Cartridge R/O - Distilled Feed
D50231
Ultrapure Mixed Bed Cartridge
D50233
Filter, 0.2 Micron
FL703X2
2.
Place the EASYpure LF on a bench top convenient to
your work area, noting the Clearance Requirements.
3.
Tape the 1/4" OD x 1/4" NPT tubing adapter using 1
1/2 to 2 turns of Teflon® tape, and install it on your in-
coming water line or your storage reservoir.
4.
Install one end of the tubing to this adapter, but do not
install the other end on the EASYpure LF until car-
tridges have been installed. You will connect the
water supply to the quick disconnect body in the Initial
Operation section. Refer to the Tubing Adapter In-
stallation section of this manual.
NOTE
Cartridges and the 0.2 micron final filter are not
provided with the EASYpure LF and must be or-
dered separately.
WARNING
Do not locate the EASYpure LF directly over
equipment that requires electrical service.
Routine maintenance of this unit may involve
water spillage and subsequent electrical shock
hazard if improperly located.
Do not use in the presence of flammable or
combustible materials; fire or explosion may
result. This device contains components which
may ignite such materials.
This device is to be used with water feeds only.
Sanitizing/cleaning agents must be used in
compliance with instructions in this manual.
Failure to comply with the above could result in
explosion and personal injury.
NOTE
If you purchased a wall mount bracket for the
EASYpure LF, refer to the Wall Mounting sec-
tion of this manual.
NOTE
The outlet of your storage reservoir must be
level with the inlet of the EASYpure LF.
NOTE
To disrupt the flow of water to the EASYpure LF,
press the stainless steel thumb pad on the
quick disconnect fitting. The insert with the
valve is easily removed. To resupply the
EASYpure LF with water, push the insert into
the coupling body and listen for the click that
indicates full closure.
Unpackaging and Installation