Introduction

The Barnstead EASYpure LF is a line-fed water purifica-
tion system designed to provide Type I reagent-grade
water with extremely low organic content. It uses a
three-stage deionization process combined with a 0.2
micron filter to polish pretreated water (distilled, deion-
ized, or reverse osmosis) to produce water with a resis-
tivity of up to 18.3 megohms-cm and with a total organic
carbon content of less than 15 ppb. Water resistivity is
continuously sensed by a resistivity cell and displayed
on a digital display.

Introduction

Figure 1 EASYpure LF Front

Figure 2 EASYpure LF Back

Summary of Contents for EASYpure LF

Page 1: ...1 BARNSTEAD THERMOLYNE CORPORATION EASYpure LF OPERATION MANUAL AND PARTS LIST Series 1052 LT1052X2 9 8 97...

Page 2: ...fications 6 Dimensions and Clearance Requirements 6 Feedwater Requirements 6 Product Water 6 Electrical Requirements 6 Environmental Conditions 6 Declaration of Conformity 7 Unpackaging and Installati...

Page 3: ...it in conformance with local electrical codes For safe operation please pay attention to the alert boxes throughout the manual Alert Boxes WARNING Warnings alert you to a possibility of personal inju...

Page 4: ...5 Disconnect from the power supply prior to maintenance and servicing To avoid personal injury 1 Do not use in the presence of flammable or combustible materials fire or explosion may result This devi...

Page 5: ...n process combined with a 0 2 micron filter to polish pretreated water distilled deion ized or reverse osmosis to produce water with a resis tivity of up to 18 3 megohms cm and with a total organic ca...

Page 6: ...0 Ohms cm Deionized 1 0 Megohms cm Reverse osmosis 100 000 Ohms cm Product Water Quality Resistivity ASTM Type I TOC Less than 15 0 ppb Flow Rate 1 5 LPM at 30 PSIG inlet feedwater pressure at 60Hz an...

Page 7: ...al equipment for measurement control and laboratory use Part I General Requirements per the provisions of the Electromagnetic Compatability Directive 89 336 EEC as amended by 92 31 EEC and 93 68 EEC a...

Page 8: ...l filter are not provided with the EASYpure LF and must be or dered separately WARNING Do not locate the EASYpure LF directly over equipment that requires electrical service Routine maintenance of thi...

Page 9: ...dapter DO NOT use the o ring at this time 4 Push the tubing through the nut until it bottoms out in the adapter 5 Remove the adapter nut and tubing Place the o ring over the tubing Be careful not to p...

Page 10: ...ed in the EASYpure LF where the disconnect body was installed 7 Remove the four feet from the EASYpure LF and retain the screws 8 Place the EASYpure LF on the wall bracket swivel base so the screw hol...

Page 11: ...urce to the quick disconnect body on the outside of the bracket cannot be done until cartridges have been installed You will connect the water supply to the quick dis connect body in the Initial Opera...

Page 12: ...12 Figure 5 Cartridge Installation WALL MOUNTING O rings Upper end cap Flanges O rings Lower end cap...

Page 13: ...ough the cartridges to drain 13 Remove a new 0 2 micron filter and bell assembly from its bag and insert it into the Luer fitting Gently turn it clockwise until it is fully seated in the Luer fit ting...

Page 14: ...re LF on a continuous basis throughout the day it is recommend ed that the EASYpure LF be left in the run mode In the run mode the purity meter display indicates the resistivity of the water available...

Page 15: ...tion in structions included with the float switch for installa tion to tank 3 If using D2706 low pressure switch install the PVC tee supplied with D2706 in incoming water line Screw the switch into th...

Page 16: ...cartridges See note below 10 Install new cartridges and rinse according to the in structions for Cartridge Installation in the Initial Op eration section WARNING Disconnect from the power supply prio...

Page 17: ...er was installed To sanitize the EASYpure LF the purification cartridges must be replaced with a sanitization cartridge and the two empty cartridges The simple to use saniti zation cartridge Catalog N...

Page 18: ...idges and rinse according to the instructions for Cartridge Installation and Rinse Up in the Initial Operation section Fuse Replacement 1 Turn the system off 2 Disconnect incoming water line at quick...

Page 19: ...cov er 6 Remove the cover by lifting it straight up 7 Remove the filter and the Luer fitting 8 Remove the screws securing the front cover Re move the front cover 9 Disconnect the membrane switch lead...

Page 20: ...circuit board relay plug 16 Reconnect the membrane switch lead 17 Reinstall the front cover 18 Retape the Luer fitting with new Teflon tape and rein stall fitting 19 Reinstall the EASYpure LF s cover...

Page 21: ...efully remove the O ring before cleaning the cell 11 Wash the cell in a mild detergent solution or a 10 Hydrochloric or Sulfuric acid solution follow acid manufacturers warnings and recommended han dl...

Page 22: ...16 Reconnect the power cord and the pump interlock plug to the unit and reconnect the unit to the power supply Connect incoming water line at quick discon nect Shutdown If the EASYpure LF is to be sh...

Page 23: ...d valves Jumper or devices not Install installed in pump interlock Pump worn out or defective Replace pump Display reads Err when Resistivity cell lead Check resistivity cell lead checking resistivity...

Page 24: ...acture Replace the cartridges with unexpired ones Change in feedwater characteristics If a Barnstead ROpure is the feedwater source check that the membrane is functioning properly If a Barnstead Still...

Page 25: ...r Barnstead Thermolyne representative for additional information on the expendables kit Description Catalog No Recommended Quantity Pretreatment Cartridge DI Feed D50230 1 Pretreatment Cartridge R O D...

Page 26: ...tock of those parts listed in the GENERAL MAINTENANCE PARTS and SAFETY STOCK sections Description Catalog No Recommended Quantity Display PCB PC703X1 1 Main PCB 100 120 Volt PC703X2 1 230 230 Volt PC7...

Page 27: ...pur chased this unit or can be obtained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer If the dealer cannot handle your request the...

Page 28: ...28 Figure 8 Flow Diagram Flow Diagram...

Page 29: ...SE DRAW WITH SWITCH AND POWER ENTRY MODULE POWER CORD MEMBRANE BLUE AND YELLOW CELL 3 BLACK MOTOR PUMP RED ORANGE MONITOR RESISTIVITY PCB S 2 1 4 SWITCH RED BLACK TRANSFORMER 100 120 VAC WHITE LINE FI...

Page 30: ...ENTRY MODULE POWER CORD OPTIONAL PRESSURE CAT NO D2706 SWITCH 2 4 3 YELLOW BLUE AND OPTIONAL FLOAT CAT NO D0603 SWITCH CR 1 ORANGE BLACK MOTOR PUMP RED 2 3 1 4 2 3 4 FOR USE WITH LINE FILTER 120 V A C...

Page 31: ...WITH SWITCH AND POWER ENTRY MODULE POWER CORD BLACK OPTIONAL PRESSURE CAT NO D2706 SWITCH 2 4 3 YELLOW OPTIONAL FLOAT CAT NO D0606 SWITCH CR 1 ORANGE BLUE MOTOR PUMP RED 2 3 1 4 2 3 4 FOR USE WITH LI...

Page 32: ...POWER CORD BLACK OPTIONAL PRESSURE CAT NO D2706 SWITCH 2 4 3 YELLOW OPTIONAL FLOAT CAT NO D0606 SWITCH CR 1 ORANGE BLUE MOTOR PUMP RED 2 3 1 4 2 3 4 FOR USE WITH LINE FILTER 230 V A C MONITOR RESISTIV...

Page 33: ...33...

Page 34: ...34...

Page 35: ...rmolyne at its own dis cretion Heating elements however because of their susceptibility to overheating and con tamination must be returned to our factory and if upon inspection it is concluded that fa...

Page 36: ...36 Barnstead Thermolyne 2555 Kerper Blvd P O Box 797 Dubuque IA 52004 0797 USA PHONE 319 556 2241 800 553 0039 FAX 319 589 0516 E Mail mkt barnsteadthermolyne com a subsidiary of...

Reviews: