45
reduce the effectiveness of the lamp. If the glass
portion of the lamp is touched, clean it with a
damp, lint-free cloth: use isopropyl alcohol as
required.
5.
Insert the lamp halfway into the UV lamp cham-
ber and plug UV cable in, then fully insert the
lamp into the chamber.
6.
Make sure the UV lamp cable is tucked behind
the chamber before closing and relatching the
door.
7.
Reconnect the power cord. Turn unit ON.
8.
Reset UV Timer according to “Resetting the UV
Timer.”
9.
Operate normally.
Fuse Replacement
1.
Turn off the NANOpure DIamond Life Science
(UV/UF) and disconnect it from the power sup-
ply by removing the power cord directly below
the main power switch on right side of unit.
Locate the fuse drawer above the power cord
receptacle.
2.
Pull out the fuse drawer located in the power
entry module.
3.
Remove old fuses and replace with
fuses of the same type and rating. (See
“Replacement Parts.”)
4.
Replace fuse drawer.
5.
Reattach the power cord.
6.
Operate normally.
M
AINTENANCE AND
S
ERVICING
Warning
Replace fuses with those of the same
type and rating.