Choosing a Site
The NANOpure DIamond Life Science (UV/UF) features a
removable control panel display which allows the system
to be mounted almost anywhere within the laboratory.
Use the wall bracket for wall mounted systems as a tem-
plate to drill mounting holes. (The NANOpure DIamond
Life Science (UV/UF) does not include screws and
fasteners for mounting.) Please refer to the “General
Specifications” section for clearance requirements.
Installation
NOTE
The outlet of a gravity feed storage
reservoir must be above or at the
same level as the inlet of the
NANOpure DIamond Life Science
(UV/UF).
********************
************************
Caution
Wall composition, condition and
construction as well as fastener type
must be considered when mounting
this unit. The mounting surface and
fasteners selected must be capable of
supporting a minimum of 150 lbs.(68
kg). Inadequate support and/or fasten-
ers may result in damage to mounting
surface and/or equipment. If you are
unsure of mounting surface
composition, condition and construc-
tion or correct fasteners, consult your
building maintenance group or
contractor.
Warning
Do not locate the NANOpure DIamond
Life Science (UV/UF) directly over
equipment that requires electrical
service. Routine maintenance of this
unit may involve water spillage and
subsequent electrical shock hazard if
improperly located.
Do not use in the presence of
flammable materials; fire or explosion
may result. This device contains
components which may ignite such
materials.