Ultrafilter Installation
Remove the ultrafilter from its protective bag. Install the
ultrafilter as follows:
1.
Open the front (right side) door and right side
door.
2.
Remove restraints (twist tie) from the three
tubes.
3.
Locate the metal bracket and separate the
Velcro
®
strap.
4.
Remove the cap on the top side port of the
ultrafilter (make sure the bottom cap on the bot-
tom side port is secure). Insert side ports into
the holes in the metal bracket.
5.
Align the ultrafilter with the metal bracket (“This
End Up” should be facing front) and push the
bottom fitting tightly into the bottom connector.
6.
Push the top fitting of the ultrafilter into the top
connector, making sure the connection is tight.
7.
Secure the ultrafilter with the velcro strap.
8.
Connect tubing with luer fitting to the top side
port.
9.
Close and latch right side door and front (right
side) door.
I
NSTALLATION
Ultrafilter Installation
Note
Do not install the ultrafilter until the
NANOpure DIamond Life Science
(UV/UF) is ready for use. If the unit
sits for longer than 72 hours without
use, it will become contaminated and
may affect product water quality.
10
Ultrafilter with fittings
Bottom side port with cap
Top side port with cap
removed
Metal bracket
Velcro strap
Ultrafilter Connections (Right side view)
Bottom connector
Top connector
Note
The ultrafilter must be installed before
the 0.2 micron filter. Please refer to
“0.2 Micron Filter Replacement” under
“Maintenance and Servicing” for
instructions on how to replace the 0.2
micron filter.
Top side port
tubing with
luer fitting
® Velcro is a registered Trademark of Velcro Industries B.V.