13
OPERATION:
(Con’t)
TIMER: (Con’t)
Conversely, if the original selection is 1.5 minutes and 0.5 minutes have
elapsed when a new timing interval of 1.0 minutes is entered, the timer will
subtract the 0.5 minutes already elapsed and show 0.5 minutes as remaining.
NOTE:
IF THE ACCUMULATED TIME REACHES 99.9, THE SHAKER WILL CONTINUE TO
SHAKE AND THE TIMER DISPLAY REMAINS AT 99.9
NOTE:
THE SHAKER DOES NOT OPERATE IF THE TIMER IS SET AT 00.0.
STARTING THE SHAKING ACTION:
Once the timing selection has been made, press the start button in the
speed control module to initiate the shaking action.
If, for any reason, shaking action needs to be stopped, press the stop
button in the speed control module.
TEMPERATURE CONTROL MODULE:
To set the desired operating temperature, use the Up and Down arrow
keys. Once set, temperature control is initiated by pressing the heat on button.
The heater will react and start increasing the temperature to reach the setpoint.
Temperature controller is PID type.
NOTE:
MAXIMUM TEMPERATURE SETTING IS 60ºC.
During operation, both the Up and Down arrow keys can be used to adjust
the temperature to a new setpoint.
SAFETY THERMOSTAT:
After the unit has stabilized at the temperature setpoint, turn the safety
thermostat counterclockwise until the hi-limit legend comes on. Then turn the
safety thermostat clockwise about 5-degrees of rotation.
NOTE:
EACH TIME THE UNIT'S TEMPERATURE IS CHANGED, THE SAFETY
THERMOSTAT
SETTING WILL HAVE TO BE CHANGED. THE UNIT IS DESIGNED FOR
TEMPERATURES
SLIGHTLY ABOVE AMBIENT TO 60ºC.
DANGER:
TO AVOID INJURY FROM BURNS, DO NOT TOUCH OR
OTHERWISE CONTACT HEATED SURFACES WHEN UNIT IS OPERATING
AT ELEVATED TEMPERATURES
DANGER:
DO NOT OPERATE THE UNIT IF ANY OF THE TEMPERATURE
CONTROLS BECOME INOPERATIVE—OTHERWISE, A HAZARDOUS
CONDITION WILL DEVELOP WHICH CAN RESULT IN INJURY OR DEATH
AND PROPERTY DAMAGE.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com