51
51
230 VAC to 115 VAC
Parts Required
•
two 15 AMP time delay fuses (Part No.
266058)
•
jumper for Power I/O Board
•
jumper for Printer Power PC Board
•
suitable 115 VAC plug or plug adapter
1.
On the Power I/O PC Board: Remove jumper
from J7 to J8. Install jumpers from J5 to J7
and from J3 to J8.
2.
Move black heater wire from J11 to J14. Move
white heater wire from J12 to J11. Leave the
other black wire connected.
3.
On the Printer Power PC Board (MC10 with
printer only): Remove jumper from JMP3 posi-
tion and install jumpers in JMP1 and JMP2
positions.
4.
Install the 15 AMP fuses in fuseholders.
5.
Reinstall the left side panel, rear access panel
and pressure relief valve deflection cover.
6.
Use either a plug adapter or remove the
power cord plug and install a suitable 115
VAC plug.
7.
Tag the sterilizer as configured for 115 VAC.
A
PPENDIX