19
Installing Overflow Tube
(Optional)
If the sterilizer is located within 5 ft. (1.5 m) of a sink or
drain, the overflow tube may be used. The collection bottle
will not require emptying while the automatic overflow is
used.
1.
Push the overflow tube (supplied) into the upper
quick connect on the back of the cooling tank.
2.
Push the tube connector in until it “clicks.”
3.
Place the end of the overflow tube in a sink or
drain(atmospheric drain).
O
PERATION
Note
The sink or drain must be below the
level of the cooling tank for the
water drain.
Warning
Volatile Reaction: Do not use this
equipment for processing volatile
substances or for any purpose other
than its intended design. Failure to
observe this warning could cause
personal injury or damage to equip-
ment.
Warning
Burn Hazard: To prevent the possi-
bility of injury from liquids boiling
over and/or containers exploding,
use only self-venting automatic seal-
ing stoppers recommended by man-
ufacturers when processing all con-
tainer liquids or media.
Caution
Be sure items can withstand cycle tem-
perature before placing them in steriliz-
er. If in doubt, refer to device manufac-
turer’s instructions.
Note
Be sure to select the correct cycle for
the type of load. Sterilization cannot
be assured if maximum load size and
density are exceeded. Total specified
weight should not be exceeded for
any sterilizer cycle.
Note
Use the LIQUIDS cycle to process
glassware.
Summary of Contents for Harvey Hydroclave MC8
Page 49: ...49 49 APPENDIX Figure 32 Power I O PC Board 230 VAC...
Page 50: ...50 50 Figure 33 MC10 Printer Power Supply Board 230 VAC APPENDIX...
Page 52: ...52 52 Figure 34 Power I O Board 115 VAC APPENDIX...
Page 53: ...53 53 APPENDIX Figure 35 MC10 Printer Power Supply Board 115 VAC...
Page 54: ...54...
Page 55: ...55...