32
32
Cleaning and Disinfecting
Exterior Surfaces
The exterior surfaces may be cleaned and disinfected with
agents typically found in dental and medical offices. Wipe
the surface clean with mild detergent and disinfect in
accordance with the instructions on the disinfectant agent
container.
Every 25 Cycles
Cleaning Reservoir
Draining Reservoir
A drain fitting and plastic tubing are provided to drain the
reservoir.
1.
Insert the open end of the plastic tubing into a
container.
2.
Insert the drain fitting on the other end of the
plastic tubing into the drain port. As soon as the
drain fitting locks into place, water will begin to
exit the drain.
3.
When the water flow stops, press the tab on the
drain port to release the drain hose and fitting.
4.
Store the drain tube and fitting.
O
PERATOR
M
AINTENANCE
Figure 23
Draining Reservoir
Warning
Water May Be Hot
Drain
Fitting
Click In
To Drain
Release
Tab
Push To
Release
Drain
Port
Figure 24
Cleaning Reservoir
Clean Reservoir With
Disinfectant
Rinse Thoroughly
Before Filling With
Distilled Water
Caution
Do not use steam to disinfect or
clean the sterilizer cabinet. Extreme
high temperatures could cause
plastic parts to become distorted.
Caution
Do not use abrasive cleansers.
Hot Surface
Burn Hazard: Be sure sterilizer is
cool before cleaning reservoir, The
water in the reservoir could be HOT
if sterilizer was in use just prior to
cleaning reservoir.
Note
Sterilizers equipped with an optional
non-recirculating system may
require less frequent reservoir clean-
ing.
Hot Surface
Burn Hazard: The water may be
HOT if a cycle has been run recent-
ly. Burns could occur if hot water
comes in contact with skin.
Summary of Contents for Harvey Hydroclave MC8
Page 49: ...49 49 APPENDIX Figure 32 Power I O PC Board 230 VAC...
Page 50: ...50 50 Figure 33 MC10 Printer Power Supply Board 230 VAC APPENDIX...
Page 52: ...52 52 Figure 34 Power I O Board 115 VAC APPENDIX...
Page 53: ...53 53 APPENDIX Figure 35 MC10 Printer Power Supply Board 115 VAC...
Page 54: ...54...
Page 55: ...55...