47
47
Voltage Conversion
(Dual Voltage Models Only)
(Spec plate must state 115/230 Volt)
1.
Remove the left side panel (4 screws), the
pressure relief valve deflection cover (MC8
Only — 1 screw), and the rear access panel
(2 screws).
2.
Remove fuses from fuseholders. Discard or
save for future reconversion.
Appendix
Figure 31
Removing access panels
Power
I/O Board
Terminal
Block
Transformer
PC Board
(MC10 Printer
Option)
Warning
Unplug power cord before removing
any access panels. Voltage conver-
sion should be performed by a quali-
fied service technician.
Note
Dual voltage models have both volt-
ages listed on the rating plate. Do
not attempt to convert a unit that
does not have two fuseholders or is
not listed as a 115V/230V model.