35
35
O
PERATOR
M
AINTENANCE
Replacing Door Gasket
1.
Open the sterilizer door and remove the hex
nut in the center of the gasket retainer and
remove the gasket retainer and gasket.
2.
Clean surface of the door plate, new gasket
and gasket retainer of any dirt, adhesives or
deposits.
3.
Apply a generous amount of silicone multipur-
pose sealant (available in most hardware
stores) in a 2” diameter area around the cen-
ter stud of the door plate.
4.
Insert the retainer inside the gasket (outer lip
of retainer against gasket) as shown.
5.
Slide the gasket and retainer assembly onto
the center stud of the door plate and secure it
finger tight (firm). Do not use wrench.
Figure 28
Door Gasket Replacement
Silicon (RTV)
Sealant
Gasket
Gasket
Retainer
Hex Nut
Hole at 4:00
position
Gasket Retainer
(Chamber Side)
Gasket
Door Side
Section Of
Door Gasket and
Gasket Retainer
Note
The gasket is adhered to the door
plate and must be pulled loose.
Note
Allow about an hour for the silicone
sealant to set, then run a no-load
cycle to seat gasket.