Additional Information
Cancel Cycle
To cancel a sterilizing cycle or drying phase in progress,
press the STOP switch.
Emergency Power Off
To remove all power from the sterilizer, press the OFF
switch, then unplug the power cord.
Power Failure During Cycle
If a power failure occurs during a cycle, the current pro-
gram will be stored in memory.
When power is restored, the controls will attempt to con-
tinue the cycle or will cancel the cycle depending on the
length of the power interruption.
If you cannot start a cycle after power is restored:
1.
Unplug the unit.
2.
Press and hold the OFF button while plugging
the unit back in.
3.
Release the OFF button after “PERFORMING
SELF-TEST” is displayed.
Warning
Volatile Reaction: Do not open the
door if there is a liquid load in the
chamber. The containers may boil
over or explode. Call for authorized
service.
Caution
The door will not open if there is
pressure or vacuum in the cham-
ber. Do not attempt to force the
door open. Press OFF, then ON, to
vent the chamber so that the door
can be opened.
Note
If a cycle is canceled prematurely,
reprocess the load. Failure to do so
may result in non-sterile goods.