Remove the reservoir lid and wipe the inside
of
reservoir with a clean damp cloth and mild
disinfectant. Do not use abrasive cleaners.
Wipe the reservoir clean of any disinfectant
residue.
6.
Remove and clean the reservoir filter.
7.
Replace the reservoir lid and refill the reser-
voir to the FILL mark in the fill funnel with dis-
tilled or deionized water.
Cleaning Reservoir Filter
1.
Unscrew the reservoir filter from its fitting and
rinse thoroughly with clean water.
2.
Scrub the outside of the filter with a stiff-bris-
tled brush (toothbrush) or clean in an ultrason-
ic cleaner.
If the above cleaning method does not effectively clean
the filter, replace it.
Cleaning Chamber Filter Screen
1.
Pull the filter screen assembly out of the water
fill port inside the chamber.
2.
Clean any deposits from the screen using a
small stiff-bristled brush (toothbrush) or ultra-
sonic cleaner.
3.
When clean, push the open end of the filter
screen into the water fill port in the back of the
chamber.
O
PERATOR
M
AINTENANCE
Figure 25
Reservoir Filter Screen
To
Remove
Note
Reservoir cleaning is not required as
often on units with non-recirculating
feature.
Pull Out
To Remove
Temperature
Probe
Filter Screen
Assembly
Water Fill
Port
Hot Surface
Burn Hazard: Be sure sterilizer is
cool before removing chamber filter
screen.
Caution
Operating the sterilizer without the
filter screen in place will damage
the unit.
Figure 26
Filter Screen
Summary of Contents for Harvey Hydroclave MC10
Page 49: ...49 49 APPENDIX Figure 32 Power I O PC Board 230 VAC ...
Page 50: ...50 50 Figure 33 MC10 Printer Power Supply Board 230 VAC APPENDIX ...
Page 52: ...52 52 Figure 34 Power I O Board 115 VAC APPENDIX ...
Page 53: ...53 53 APPENDIX Figure 35 MC10 Printer Power Supply Board 115 VAC ...
Page 54: ...54 ...
Page 55: ...55 ...