22
Instructions For Use
Absent ECG signal
ECG Screen (main/reference screen)
is not visible
6.3 Magnetic Navigation Troubleshooting
Magnetic navigation does not
detect stylet
Note: Depth icons will
not be present.
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Sensor / sensor cables damaged.
Call technical hotline at (800) 443-3385 to replace sensor and/or cables.
Sensor disconnected.
Ensure the sensor is properly connected.
Sensor is not communicating with display.
Reboot the display.
Incompatible sensor.
Use appropriate sensor. Contact customer service (800) 545-0890.
Incorrect PICC compatibility is selected.
Make sure to have [
] selected from the menu when using a catheter kit labeled
[
].
Sensor incorrectly oriented.
Orient sensor correctly per procedure.
Stylet outside sensor depth range (3 to 11 cm).
- If the vein is deep, ensure that the top of the sensor on the insertion side is touching
the patient's skin (e.g. by using tape).
- If the vein is shallow, lift the sensor away from the skin (e.g., by using a folded blanket).
Too much time has passed since calibration. Magnetic fields have changed.
Pull catheter tip back at least 12 inches (30 cm) from the sensor, recalibrate and
re-advance catheter.
Stylet did not advance under the sensor.
Pull catheter back and re-advance.
Stylet not at the catheter tip.
Ensure stylet tip is within 1 cm of distal end of catheter. Adjust as necessary.
Incompatible catheter.
Verify that the Bard Access Systems’ catheter kit has a [
] icon or [
] icon.
Stylet magnet trimmed off.
Replace catheter with a new Bard Access Systems’ catheter from a kit labeled
[
] or [
] .
Stylet is out of sensor's depth range (3 to 11 cm)
1. Pull catheter back until stylet icon is stable.
2. Verify catheter is moving in the correct direction and continue placement using
ECG guidance and catheter measurement.
Magnetic interference.
Pull catheter tip back at least 12 inches (30 cm) from the sensor, recalibrate and
re-advance catheter.
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Summary of Contents for Sherlock 3CG TCS
Page 2: ...For use with Site Rite 6 Ultrasound System Instructions for Use...
Page 9: ...7 Sherlock 3CG Tip Confirmation System 3 2 Sherlock 3CG TCS Graphical Interface...
Page 10: ...8 Instructions For Use...
Page 11: ...9 Sherlock 3CG Tip Confirmation System...
Page 28: ...26 Instructions For Use...