20
Instructions For Use
Magnet Error
Battery Error
Printer Error
Storage Device Error
Storage Device Error
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Interference caused by magnetic field changes or by sensor movement.
1. Move all metal objects at least 12 inches (30 cm) away from the sensor.
2. Lower bed rails.
3. Move all active electronic devices at least 5 feet (1.5m) away.
4. Do not move the sensor and ask the patient to remain still.
5. Ensure catheter tip is at least 12 inches (30 cm) away from the sensor.
6. Calibrate.
7. If the error continues, discontinue use and call the technical support
hotline at (800) 443-3385.
Battery malfunction.
Call the technical support hotline at (800) 443-3385.
Printer error.
1. Check and refill paper.
2. Refer to Printer Instructions for Use.
Storage device error.
Replace external USB storage device.
Storage device error.
External USB storage device has no storage space available.
1. Remove files from external USB storage device to have available storage space.
2. Replace external USB storage device.