Bard Sherlock 3CG TCS Instructions For Use Manual Download Page 20

 

18

Instructions For Use

Step 5: Position Sensor 

- Remove the adhesive backing from the sensor holder and place the sensor directly on the patient's skin with the adhesive side down. Place the sensor as flat  

   as possible for best results.
Note:

 The sensor should be positioned the same for left or right side placements.

Step 6:  Perform Initial Magnet Navigation Calibration

- Calibrate the Sherlock 3CG™ TCS by selecting CALIBRATE [  

  ] prior to setting up the sterile field to ensure there is no environmental interference.

Tip:

 If calibration fails, remove any items that may be causing magnetic interference (e.g. active motor driven equipment, monitor leads, cell phones,  

       name tags, jewelry, etc.).

Step 7: Prepare Catheter Sterile Field

Refer to the Bard Access Systems’ catheter Instructions for Use.

Step 8: Access the Vein

- Refer to the Bard Access Systems’ Site~Rite™ 6 Ultrasound System Instructions for Use and Bard Access Systems’ catheter Instructions for Use.

Step 9: Perform Final Magnet Navigation Calibration

- Ensure the catheter tip is at least 12 inches (30 cm) away from the sensor before calibrating.
- Select CALIBRATE [ 

 ] immediately prior to catheter insertion.

- Once calibration is complete, ask the patient to remain still and do not reposition the patient.
- Refer to Bard Access Systems' catheter Instructions for Use for catheter insertion.

Step 10: Insert Catheter

- Refer to Bard Access Systems’ catheter Instructions for Use for catheter insertion.

Step 11: Catheter Tip Guidance and Positioning

- Refer to Bard Access Systems’ catheter Instructions for Use for catheter insertion.
- Initially a searching magnifying glass will indicate that the stylet tip is outside the sensor range.
- Use a slow steady motion while advancing the catheter.
Magnetic Navigation
- As the stylet tip approaches the sensor, an icon appears at the edge of the screen indicating the approach of the stylet tip.
- When the stylet is under the sensor, the stylet and depth icons will display the location, orientation, and depth of the stylet in relation to the sensor.
- Advance the catheter slowly to achieve optimal performance (1 cm per second). There may be a slight delay from the time the catheter is moved until  

  the stylet icon moves on the display. Advancing the catheter too quickly may result in erratic movements of the stylet icon on the display.
- Insert the catheter until the magnetic navigation shows the stylet moving consistently downward.
- Continue to slowly advance catheter until the catheter is inserted to the external measurements as determined in the Bard Access Systems’ catheter  

  instructions for use.

Summary of Contents for Sherlock 3CG TCS

Page 1: ...on date March 2014 Bard Sherlock Sherlock 3CG and Site Rite are trademarks and or registered trademarks of C R Bard Inc All other trademarks are the property of their respective owners 2014 C R Bard I...

Page 2: ...For use with Site Rite 6 Ultrasound System Instructions for Use...

Page 3: ...Operating Temperature Parameters Do not dispose with ordinary municipal waste Operating Humidity Parameters Storage Humidity Parameters unpackaged Storage Humidity Parameters packaged Keep Dry Manufa...

Page 4: ...Tracking Calibration Step 8 Prepare Catheter Sterile Field Step 9 Access the Vein Step 10 Attach Catheter Stylet to Fin Assembly Step 11 Perform Final Magnet Navigation Calibration Step 12 Insert Cat...

Page 5: ...y positioning the tip of catheters in adult patients The primary endpoints were to assess the performance of the Sapiens technology with respect to 1 the accuracy of the Sapiens technology with regard...

Page 6: ...m controls for using Sherlock 3CG TCS Mode on the Site Rite 6 Ultrasound System Authorized accessories include Site Rite 6 Ultrasound System Sherlock 3CG TCS Sensor Sherlock 3CG TPS Stylet included wi...

Page 7: ...MRI devices The high magnetic fields created by an MRI device will attract the equipment with a force sufficient to cause death or serious injury to persons between the equipment and the MRI device T...

Page 8: ...e roll stand see illustrations below Caution When the sensor is not in use store in the holster roll stand basket or other secure location to avoid damage 2 2 Connecting the Sherlock 3CG TCS Sensor to...

Page 9: ...7 Sherlock 3CG Tip Confirmation System 3 2 Sherlock 3CG TCS Graphical Interface...

Page 10: ...8 Instructions For Use...

Page 11: ...9 Sherlock 3CG Tip Confirmation System...

Page 12: ...dio On Select to mute the magnet navigation audio Audio Mute Select to turn the magnet navigation audio on Freeze Select to copy the current ECG waveforms from the Main Screen to the Reference Screen...

Page 13: ...nu Window Patient Information Window Settings Window or quit Demonstration Mode Exit Site Marking Controls Increase Select to cycle through the exit site marking options in a positive progression Decr...

Page 14: ...he Menu Window Settings Controls P wave Highlighting Select to enable or disable highlighting the P wave on the ECG signals ECG Color Select to enable or disable the intravascular yellow ECG waveforms...

Page 15: ...ignals are only applicable to the magnetic navigation capability of the device Tone 4 is applicable to ECG in patients where there are no alterations to the cardiac rhythm that change the presentation...

Page 16: ...Rite 6 Ultrasound System via USB cable Verify that the ECG flatline signal is scrolling Verify the battery charge is sufficient for the procedure Enter Patient Information as needed Step 2 Position P...

Page 17: ...r to the umbilicus and laterally along the mid axillary line Caution Placement of red electrode outside of this region may result in reduced ECG performance Warning Place skin electrodes carefully at...

Page 18: ...on As the stylet tip approaches the sensor an icon appears at the edge of the screen indicating the approach of the stylet tip When the stylet is under the sensor the stylet and depth icons will displ...

Page 19: ...nnect the sensor to the Site Rite 6 Ultrasound System via USB cord Verify the battery charge is sufficient for the procedure Enter patient Information as needed Step 2 Position Patient and Perform Ult...

Page 20: ...patient to remain still and do not reposition the patient Refer to Bard Access Systems catheter Instructions for Use for catheter insertion Step 10 Insert Catheter Refer to Bard Access Systems cathet...

Page 21: ...ectrical outlet to run the system on battery power Sensor cannot be detected by the display Call the technical support hotline at 800 443 3385 Interference caused by magnetic field changes or by senso...

Page 22: ...sor and ask the patient to remain still 5 Ensure catheter tip is at least 12 inches 30 cm away from the sensor 6 Calibrate 7 If the error continues discontinue use and call the technical support hotli...

Page 23: ...y Patient movement Ask the patient to remain still Ensure the patient is warm and relaxed Extraneous noise Unplug the Site Rite 6 Ultrasound System from AC power Make Sure the Ultrasound probe is not...

Page 24: ...y using tape If the vein is shallow lift the sensor away from the skin e g by using a folded blanket Too much time has passed since calibration Magnetic fields have changed Pull catheter tip back at l...

Page 25: ...se Solution Cause Solution Cause Solution Cause Solution Stylet at edge of sensor s depth range 3 to 11 cm 1 Pull catheter back until stylet icon is stable 2 Verify catheter is moving in the correct d...

Page 26: ...llowing actions void the warranty of the Sherlock 3CG TCS Sensor and may result in injury or equipment damage Opening or servicing the Sherlock 3CG TCS Sensor by anyone other than Bard Access Systems...

Page 27: ...ures may result in damage to the system Warning Opening or servicing the Sherlock 3CG TCS Sensor by anyone other than Bard Access Systems authorized service personnel will void the warranty and may re...

Page 28: ...26 Instructions For Use...

Reviews: