19
Sherlock 3CG™ Tip Confirmation System
- Below is an image of Magnetic Navigation only mode.
6 Trouble Shooting and Error Messages
6.1 Error Screens
Sensor Disconnected
Sensor Error
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Cause:
Solution:
Sensor not connected to the display.
Ensure the sensor is properly connected. Disconnect and reconnect the sensor.
The display power supply is connected to a noisy electrical outlet.
Disconnect the display power supply from the electrical outlet to run the system
on battery power.
Sensor cannot be detected by the display.
Call the technical support hotline at(800) 443-3385.
Interference caused by magnetic field changes or by sensor movement.
1. Move all metal objects at least 12 inches (30 cm) away from the sensor.
2. Lower bed rails.
3. Move all active electronic devices at least 5 feet (1.5 m) away.
4. Do not move the sensor and ask the patient to remain still.
5. Calibrate.
6. If the error continues, discontinue use and call the technical support
hotline at (800) 443-3385.
Sensor malfunction.
1. Turn off and restart the system.
2. If the error continues, discontinue use and call the technical support
hotline at (800) 443-3385.