
To adjust the high temperature alarm setpoint:
1. Press MENU key to go to the Main Menu screen.
2. Use UP or DOWN keys and ENTER key to enter
USER password 2000.
3. Press UP or DOWN keys to scroll to
Sys Config
;
press ENTER key.
4. Press UP or DOWN keys to scroll to
Alarm
Setpoints (A4)
.
5. Press ENTER key to scroll to
High Temp
(see
Figure 2.1).
6. Press UP or DOWN keys to adjust setpoint.
The trigger type for the low temperature, high
temperature and high temperature 2 alarms can be
changed to reference the displayed average or the
lowest value (for low temperature alarm) and the
highest value (for high temperature alarms).
To adjust the low temperature alarm trigger type:
1. Press MENU key to go to the Main Menu screen.
2. Use UP or DOWN keys and ENTER key to enter
USER password 2000.
3. Press UP or DOWN keys to scroll to
Sys Config
;
press ENTER key.
4. Press UP or DOWN keys to scroll to
Alarm
Setpoints (A4)
.
5. Press ENTER key to scroll to
Trigger Type
(see
Figure 2.1).
6. Press UP or DOWN keys to change from
Hi/Low
to
Average
.
A delay of 10 seconds is applied to each alarm. This
can be adjusted by:
1. Press MENU key to go to the Main Menu screen.
2. Use UP or DOWN keys and ENTER key to enter
USER password 2000.
3. Press UP or DOWN keys to scroll to
Sys Config
;
press ENTER key.
4. Press UP or DOWN keys to scroll to
Alarm
Setpoints (A4)
.
5. Press ENTER key to scroll to
Alarm Delay
(see
Figure 2.1).
6. Press UP or DOWN keys to adjust the delay.
High Temperature 2 Alarm
The LV will indicate a high temperature 2 alarm when
any of the connected sensors that are enabled read a
value above the high temperature 2 limit of 90°F. This
alarm
does not
use the average of the sensors like other
functions do. In addition to the alarm being generated,
this event will put the system into emergency mode.
See emergency mode for more information.
To adjust the high temperature 2 alarm setpoint:
1. Press MENU key to go to the Main Menu screen.
2. Use UP or DOWN keys and ENTER key to enter
USER password 2000.
3. Press UP or DOWN keys to scroll to
Sys Config
;
press ENTER key.
4. Press UP or DOWN keys to scroll to
Alarm
Setpoints (A4)
.
5. Press ENTER key to scroll to
High Temp
2
(see
Figure 2.1).
6. Press UP or DOWN keys to adjust setpoint.
The trigger type for the low temperature, high
temperature and high temperature 2 alarms can be
changed to reference the displayed average or the
lowest value (for low temperature alarm) and the
highest value (for high temperature alarms).
To adjust the low temperature alarm trigger type:
1. Press MENU key to go to the Main Menu screen.
2. Use UP or DOWN keys and ENTER key to enter
USER password 2000.
3. Press UP or DOWN keys to scroll to
Sys Config
;
press ENTER key.
4. Press UP or DOWN keys to scroll to
Alarm
Setpoints (A4)
.
5. Press ENTER key to scroll to
Trigger Type
(see
Figure 2.1).
6. Press UP or DOWN keys to change from
Hi/Low
to
Average
.
A delay of 10 seconds is applied to each alarm. This
can be adjusted by:
1. Press MENU key to go to the Main Menu screen.
2. Use UP or DOWN keys and ENTER key to enter
USER password 2000.
3. Press UP or DOWN keys to scroll to
Sys Config
;
press ENTER key.
4. Press UP or DOWN keys to scroll to
Alarm
Setpoints (A4)
.
5. Press ENTER key to scroll to
Alarm Delay
(see
Figure 2.1).
6. Press UP or DOWN keys to adjust the delay.
Smoke Alarm
The LV will indicate a smoke alarm when the smoke
detector input indicates there is smoke present in the
shelter. This input is enabled by default but comes
from the factory with a jumper. To utilize this input,
remove the jumper and connect the sensor in place
of the jumper. The alarm will communicate this
information to all of the wall units. The wall units will
be disabled so that no operations occur. See smoke
Page
27 of 36