6. Important information
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment –
guidance
of the transmitter.
Recommended
separation distance
d = 1.2
√
P
d = 1.2
√
P
80 MHz to 800
MHz
d = 2.3
√
P
800 MHz to 2.5
Ghz
Where P is the maximum
output power rating
of the transmitter in
watts (W) according
to the transmitter
manufacturer and d
is the recommended
separation distance in
meters (m).
Field strengths from
fi
xed RF transmitters,
as determined by an
electromagnetic site
survey,
8
should be less
than the compliance
level in each frequency
range.
9
Interference may occur in
the vicinity of equipment
marked with symbol:
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and re
fl
ection from structures, objects and people.
Recommended separation distance
The Nio Color 5MP is intended for use in an electromagnetic environment in which radiated RF distur-
bances are controlled. The customer or the user of the Nio Color 5MP can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equip-
8. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Nio Color 5MP is used exceeds the applicable RF compliance level above, the Nio Color 5MP should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the Nio Color 5MP.
9. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
K5903146 NIO COLOR 5MP 24/05/2017
39
Summary of Contents for Nio Color 5MP MDNC-6121
Page 1: ...Nio Color 5MP User Guide 5 8MP high bright color display MDNC 6121 K5903146 01 24 05 2017 ...
Page 8: ...1 Welcome 6 K5903146 NIO COLOR 5MP 24 05 2017 ...
Page 18: ...3 Daily operation Cancel Standby IEC 60417 5009 16 K5903146 NIO COLOR 5MP 24 05 2017 ...
Page 32: ...4 Advanced operation 30 K5903146 NIO COLOR 5MP 24 05 2017 ...
Page 34: ...5 Cleaning your display 32 K5903146 NIO COLOR 5MP 24 05 2017 ...