6. Important information
•
Reorient or relocate the receiving antenna or equipment.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The Nio Color 5MP is intended for use in the electromagnetic environment speci
fi
ed below. The customer
or the user of the Nio Color 5MP should assure that it is used in such an environment.
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment –
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood,
concrete or ceramic tile.
If
fl
oors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 2kV for power supply
lines
± 1kV for input/ output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
< 5% U
T
6
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
< 5% U
T
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
Mains power quality
should by that of a typical
commercial or hospital
environment. If the
user of the Nio Color
5MP requires continued
operation during power
mains interruptions, it
is recommended that
the Nio Color 5MP
be powered from an
uninterruptible power
supply or a battery.
Power frequency (47-63
Hz) magnetic
fi
eld
IEC 61000-4-8
3 A/m
Not applicable
7
Power frequency
magnetic
fi
elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Nio Color
5MP, including cables,
than the recommended
separation distance
calculated from the
equation applicable
to the frequency
6. is the a.c. mains voltage prior to application of the test level.
7. Nio Color 5MP doesn’t contain susceptible components to magnetic fields
38
K5903146 NIO COLOR 5MP 24/05/2017
Summary of Contents for Nio Color 5MP MDNC-6121
Page 1: ...Nio Color 5MP User Guide 5 8MP high bright color display MDNC 6121 K5903146 01 24 05 2017 ...
Page 8: ...1 Welcome 6 K5903146 NIO COLOR 5MP 24 05 2017 ...
Page 18: ...3 Daily operation Cancel Standby IEC 60417 5009 16 K5903146 NIO COLOR 5MP 24 05 2017 ...
Page 32: ...4 Advanced operation 30 K5903146 NIO COLOR 5MP 24 05 2017 ...
Page 34: ...5 Cleaning your display 32 K5903146 NIO COLOR 5MP 24 05 2017 ...