4. Advanced operation
4.15.4 DICOM calibration
About DICOM calibration
The DICOM calibration will add a correction to the current DICOM curve to approach the perfect DICOM
curve as well as possible.
To start DICOM calibration:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > Calibration > Embedded QA
menu.
3. Select
DICOM calibration
to start the calibration.
Warning:
Pressing a key during calibration will abort the calibration, previous values will be restored.
Note:
After calibration, the compliance check will start automatically.
4.15.5 Reset DICOM calibration
About reset DICOM calibration
It is possible to restore the original (not corrected) DICOM curve.
To reset the DICOM calibration:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > Calibration > Embedded QA
menu.
3. Enter the
DICOM preferences
submenu.
4. Select
reset DICOM calibration
to restore the original (not corrected) DICOM curve.
4.15.6 DICOM error threshold
About DICOM error threshold
The threshold to de
fi
ne the DICOM compliance can be modi
fi
ed in steps of 5% starting from 5 to 30%.
When the maximum deviation is not bigger than the selected threshold, the compliance check will be OK.
To set the DICOM error threshold:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > Calibration > Embedded QA
menu.
3. Enter the
DICOM preferences
submenu.
4. Set
DICOM error threshold
as desired and con
fi
rm.
4.16 Display orientation
About Display orientation
Your display automatically detects its physical orientation (portrait or landscape) and, by default, automat-
ically adjusts the image orientation to this. This means that when your display is physically rotated, the
image will rotate along.
The OSD menu however, allows to overrule this behavior and force the image orientation to portrait or
landscape regardless of the physical orientation of the display. This may be especially useful when oper-
ating your display with the screen facing upwards.
To set the Display orientation
1. Bring up the OSD main menu.
K5903146 NIO COLOR 5MP 24/05/2017
25
Summary of Contents for Nio Color 5MP MDNC-6121
Page 1: ...Nio Color 5MP User Guide 5 8MP high bright color display MDNC 6121 K5903146 01 24 05 2017 ...
Page 8: ...1 Welcome 6 K5903146 NIO COLOR 5MP 24 05 2017 ...
Page 18: ...3 Daily operation Cancel Standby IEC 60417 5009 16 K5903146 NIO COLOR 5MP 24 05 2017 ...
Page 32: ...4 Advanced operation 30 K5903146 NIO COLOR 5MP 24 05 2017 ...
Page 34: ...5 Cleaning your display 32 K5903146 NIO COLOR 5MP 24 05 2017 ...