K5902142 /07
Nio 5.8MP Gray
46
Electromagnetic immunity
The Nio 5.8MP Gray is intended for use in the electromagnetic environment specified below. The customer or
the user of the Nio 5.8MP Gray should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2 test levels Compliance level
Electromagnetic
environment – guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Nio 5.8MP Gray
requires continued
operation during power
mains interruptions, it is
recommended that the Nio
5.8MP Gray be powered
from an uninterruptible
power supply or a battery
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
Not applicable
4
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
Conducted RF
IEC 61000-4-6
3 Vrms (6 Vrms in ISM
bands)
150 kHz to 80 MHz
3 Vrms (6 Vrms in ISM
bands)
-
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
Immunity to RF wireless communications equipment
Test
frequency
(MHz)
Band (MHz)
Service
Modulation
Maximum
power (W)
Distance
(m)
Immunity
test level (V/
m)
385
380 – 390
TETRA 400
Pulse
modulation
18 Hz
1.8
0.3
27
450
430 – 470
GMRS 460,
FRS 460
FM ± 5 kHz
deviation
2
0.3
28
4.
Nio 5.8MP Gray doesn’t contain components that are susceptible to magnetic fields.
Summary of Contents for Nio 5.8MP Gray
Page 5: ...5 K5902142 07 Nio 5 8MP Gray Welcome 1...
Page 8: ...K5902142 07 Nio 5 8MP Gray 8 Welcome...
Page 9: ...9 K5902142 07 Nio 5 8MP Gray Installation 2...
Page 16: ...K5902142 07 Nio 5 8MP Gray 16 Installation...
Page 17: ...17 K5902142 07 Nio 5 8MP Gray Daily operation 3...
Page 22: ...K5902142 07 Nio 5 8MP Gray 22 Daily operation...
Page 36: ...K5902142 07 Nio 5 8MP Gray 36 Advanced operation...
Page 37: ...37 K5902142 07 Nio 5 8MP Gray Cleaning the display 5...
Page 39: ...39 K5902142 07 Nio 5 8MP Gray Important information 6...
Page 57: ......
Page 58: ...www barco com K5902142 07 2022 12 13 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...