K5902094 (451920611693) /02
MDSC-8232
60
White point
Native: Min 4900K, Typical 6500K, Max 9200K
Calibrated: 5600K, 6500K, 7600K, 9300K
Gamma
Native, 1.8, 2.2, 2.4, DICOM, Video
Backlight sensor
Backlight stabilization
Front sensor
N/A
Housing color
RAL 9004
Screen protection
Anti-reflective tempered glass
Keyboard
Capacitive touch keyboard (with enable/disable switch)
Video input signals
1 x 4K-UHD input: 2x DP 1.1 (at 1920x2160 or 2048x2160) genlocked or 1x
DP 1.2 MST or 2x 10Gb Ethernet fiber optic
1 x DVI FHD input (upscaled to UHD)
1 x 3G-SDI FHD input (upscaled to UHD)
Video output signals
1 x 3G-SDI loopthrough
1 x DVI FHD output (4K is down-scaled to FHD)
Video formats
Display Port 1.2 MST (10 bit) up to 4096x2160 @60Hz (displayed at native
resolution)
Dual stream DP 1.1 (10-bit) up to 2048x2160 x2 @60Hz synchronized
(displayed at native resolution)
3G-SDI (10-bit), in compliance with: SMPTE 425M (Level A), SMPTE 424M,
SMPTE 292M, SMPTE 259M, SMPTE 296M, ITU-R BT.656, ITU-R BT.601
DVI (8bit) up to 1920x1080 @60Hz
Multi-image support
No
Remote control
FW download & control protocol via optical Ethernet intferface
Power source requirements for
external power supply (nominal)
100-240 VAC, 50/60 Hz auto-switch
Power source requirements for
display power input
+24 VDC ±10% / 9.5 A
Power supply DC output cable
length
2.0 m (6.6 ft)
Power consumption
150 W (typical), 230 W (max)
Cooling
Thermo regulated forced cooling
Power management
Low power modes:
Stand-by: about 52 W
Power-off: about 1 W
OSD languages
English, French, German, Spanish, Italian
Dimensions display (W x H x D) 777 x 472 x 117 mm (30.6 x 18.6 x 4.6 in)
Dimensions power supply (W x
H x D)
223 x 88.5 x 38.6 mm (8.8 x 3.5 x 1.5 in)
Dimensions packaged (W x H x
D)
730 x 560 x 170 mm (28.74 x 22.05 x 6.69 in)
Net weight display
15.5 kg (34.1 lbs)
Net weight power supply
1.2 kg (2.65 lbs)
Net weight packaged
21.7 kg (47.9 lbs)
Mounting standard
VESA (100 x 100 mm, 200 x 100 mm)
Recommended modalities
Endoscopy, Laparoscopy
Certifications
•
IEC 60601-1: 2012 Edition 3.1 (Medical electrical equipment
–
Part 1:
General requirements for basic safety and essential performance)
•
EN 60601-1: 2006 +A1:2013 (Medical electrical equipment
”
. Part 1:
General requirements basic safety and essential performance)
•
ANSI/AAMI ES 60601-1: 2005/(R)2012 and A1:2012, C1:2009/(R)2012
and A2:2010/(R)2012 - Med. El. Equip., Part 1: general req. for basic safety
and essential performance.
•
CAN/CSA-C22.2 No. 60601-1: 14 Medical Electrical Equipment - Part 1:
General Requirements for Basic Safety and Essential Performance
(Harmonized with Ed. 3.1)
Summary of Contents for MDSC-8232
Page 5: ...5 K5902094 451920611693 02 MDSC 8232 Welcome 1 ...
Page 8: ...K5902094 451920611693 02 MDSC 8232 8 Welcome ...
Page 9: ...9 K5902094 451920611693 02 MDSC 8232 Parts controls and connectors 2 ...
Page 15: ...15 K5902094 451920611693 02 MDSC 8232 Display installation 3 ...
Page 22: ...K5902094 451920611693 02 MDSC 8232 22 Display installation ...
Page 23: ...23 K5902094 451920611693 02 MDSC 8232 Daily operation 4 ...
Page 29: ...29 K5902094 451920611693 02 MDSC 8232 Advanced operation 5 ...
Page 42: ...K5902094 451920611693 02 MDSC 8232 42 Advanced operation ...
Page 43: ...43 K5902094 451920611693 02 MDSC 8232 Important information 6 ...
Page 63: ...K5902094 451920611693 02 MDSC 8232 63 Important information ...