K5902094 (451920611693) /02
MDSC-8232
51
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
•
Reorient or relocate the receiving antenna or equipment.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The MDSC-8232 is intended for use in the electromagnetic environment (IEC 60601-1-2 4
th
edition) specified
below. The customer or the user of the MDSC-8232 should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2 4
th
edition
(2014)
Test levels
Compliance level
Electromagnetic
environment
–
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8kV contact
± 15kV air
± 8kV contact
± 15kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 2kV for power supply
lines
± 1kV for input/ output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
< 5% U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
Mains power quality
should by that of a typical
commercial or hospital
environment. If the user of
the MDSC-8232 requires
continued operation
during power mains
interruptions, it is
recommended that the
MDSC-8232 be powered
from an uninterruptible
power supply or a battery.
1:
is the a.c. mains voltage prior to application of the test level.
Summary of Contents for MDSC-8232
Page 5: ...5 K5902094 451920611693 02 MDSC 8232 Welcome 1 ...
Page 8: ...K5902094 451920611693 02 MDSC 8232 8 Welcome ...
Page 9: ...9 K5902094 451920611693 02 MDSC 8232 Parts controls and connectors 2 ...
Page 15: ...15 K5902094 451920611693 02 MDSC 8232 Display installation 3 ...
Page 22: ...K5902094 451920611693 02 MDSC 8232 22 Display installation ...
Page 23: ...23 K5902094 451920611693 02 MDSC 8232 Daily operation 4 ...
Page 29: ...29 K5902094 451920611693 02 MDSC 8232 Advanced operation 5 ...
Page 42: ...K5902094 451920611693 02 MDSC 8232 42 Advanced operation ...
Page 43: ...43 K5902094 451920611693 02 MDSC 8232 Important information 6 ...
Page 63: ...K5902094 451920611693 02 MDSC 8232 63 Important information ...