8. Important information
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment –
guidance
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
< 5% U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
< 5% U
T
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
Mains power quality
should by that of a typical
commercial or hospital
environment. If the user
of the Coronis Fusion
10MP requires continued
operation during power
mains interruptions, it is
recommended that the
Coronis Fusion 10MP
be powered from an
uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic
fi
eld
IEC 61000-4-8
3 A/m
Not applicable
Power frequency
magnetic
fi
elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Coronis Fusion
10MP, including cables,
than the recommended
separation distance
calculated from the
equation applicable
to the frequency
of the transmitter.
Recommended
separation distance
d = 1.2
√
P
d = 1.2
√
P
80 MHz to 800
MHz
d = 2.3
√
P
800 MHz to 2.5
Ghz
Where P is the maximum
output power rating
of the transmitter in
watts (W) according
to the transmitter
manufacturer and d
is the recommended
separation distance in
meters (m).
Field strengths from
fi
xed RF transmitters,
as determined by an
electromagnetic site
survey,
2
should be less
1. is the a.c. mains voltage prior to application of the test level.
K5902031 CORONIS FUSION 10MP 28/08/2013
43
Summary of Contents for MDCG-10130
Page 1: ...User Guide MDCG 10130 Coronis Fusion 10MP...
Page 2: ......
Page 3: ...Coronis Fusion 10MP User Guide MDCG 10130 K5902031 03 28 08 2013...
Page 8: ...1 Welcome 4 K5902031 CORONIS FUSION 10MP 28 08 2013...
Page 38: ...6 Repackaging instructions 34 K5902031 CORONIS FUSION 10MP 28 08 2013...
Page 40: ...7 Cleaning your display 36 K5902031 CORONIS FUSION 10MP 28 08 2013...
Page 54: ......
Page 55: ......