R5911227 /05
Coronis Fusion 4MP / 6MP
28
•
Color space:
sRGB. See
for more info.
•
Display function:
sRGB. See
for more info.
To enable sRGB:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration
menu.
3.
Select
sRGB
and confirm.
4.15 Ambient Light Compensation (ALC)
About ALC
Ambient Light Compensation (ALC) can only be enabled on your display when the display function
is set to DICOM. Please refer to
.
When ALC is enabled, the DICOM display function will be recalculated taking a preset ambient light correction
value into account. This value is determined by the selected reading room. Therefore, it is also important to
select a realistic reading room when enabling ALC. This can be done by following the instructions in
To enable/disable ALC:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Ambient Light
menu.
3.
Enter the
Ambient Light Compensation
submenu.
4.
Select
Enabled/Disabled
as desired and confirm.
4.16 Reading rooms
About reading rooms
Reading rooms can only be selected on your display when the display function is set to DICOM.
Please refer to
The American Association of Physicists in Medicine (AAPM) composed a list of pre-defined reading rooms.
Each of these reading rooms are defined by following parameters:
•
the maximum light allowed in this type of room
•
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset ambient light correction value to take
into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is enabled.
Please refer to
“Ambient Light Compensation (ALC)”, page 28
to enable ALC.
The available reading rooms for your Coronis Fusion 4MP / 6MP are:
•
CR/DR/ MAMMO:
Corresponds to light conditions in diagnostic reading rooms for computed radiology,
digital radiology or mammography. This setting has the lowest maximum ambient light.
•
CT/MR/NM:
Corresponds to light conditions in diagnostic reading rooms for computed tomography,
magnetic resonance or nuclear medicine scans.
•
Staff Office:
Corresponds to light conditions in office rooms.
•
Clinical Viewing Room:
Corresponds to light conditions in diagnostic reading rooms for clinical viewing.
•
Emergency Room:
Corresponds to light conditions in emergency rooms.
•
Operating Room:
Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
Summary of Contents for MDCC-4430
Page 1: ...ENABLING BRIGHT OUTCOMES MDCC 4430 NC MDCC 6530 NC User guide Coronis Fusion 4MP 6MP...
Page 5: ...5 R5911227 05 Coronis Fusion 4MP 6MP Welcome 1...
Page 8: ...R5911227 05 Coronis Fusion 4MP 6MP 8 Welcome...
Page 9: ...9 R5911227 05 Coronis Fusion 4MP 6MP Installation 2...
Page 16: ...R5911227 05 Coronis Fusion 4MP 6MP 16 Installation...
Page 17: ...17 R5911227 05 Coronis Fusion 4MP 6MP Daily operation 3...
Page 35: ...35 R5911227 05 Coronis Fusion 4MP 6MP Cleaning your display 5...
Page 37: ...37 R5911227 05 Coronis Fusion 4MP 6MP Repackaging instructions 6...
Page 41: ...41 R5911227 05 Coronis Fusion 4MP 6MP Image 6 7 Repackaging instructions...
Page 42: ...R5911227 05 Coronis Fusion 4MP 6MP 42 Repackaging instructions...
Page 43: ...43 R5911227 05 Coronis Fusion 4MP 6MP Important information 7...
Page 62: ...R5911227 05 Coronis Fusion 4MP 6MP 62 Important information...
Page 63: ......
Page 64: ...www barco com R5911227 05 2022 01 12 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...