R5911227 /05
Coronis Fusion 4MP / 6MP
30
•
Maximum error:
Shows the maximum error of the current DICOM curve. This is the deviation compared to
a perfect DICOM.
•
Error threshold:
Shows the error threshold. This is the maximum error allowed before a DICOM
calibration is required.
•
Time elapsed since latest compliance check:
Shows the backlight runtime since last compliance check.
•
Display Function:
Shows the current display function.
•
Ambient light compensation:
Shows the ambient light compensation status.
•
Reading Room:
Shows the selected reading room.
•
Luminance:
Shows the measured luminance.
•
Black luminance:
Shows the measured black luminance.
DICOM Calibration Status
•
No calibration executed yet:
No other information is visible
•
Calibration executed:
When the calibration is executed, the following extra information is shown: Time
elapsed since latest calibration, Display Function, Ambient Light Compensation and Reading Room.
Current DICOM Settings
•
Display Function:
Shows the current display function.
•
Ambient Light Compensation:
Shows the ambient light compensation status.
•
Reading room:
Shows the selected reading room.
To retrieve the DICOM status report:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM Status Report
to make the information visible on the screen.
4.18.3 DICOM compliance check
About DICOM compliance check
The DICOM compliance check will measure the DICOM curve of your display in different steps. After
measurement, the DICOM status report is shown.
To start DICOM compliance check:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM Compliance Check
to start the compliance check.
Warning:
Pressing a key during the compliance check will abort the check.
4.18.4 DICOM calibration
About DICOM calibration
The DICOM calibration will add a correction to the current DICOM curve to approach the perfect DICOM curve
as well as possible.
To start DICOM calibration:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM calibration
to start the calibration.
Warning:
Pressing a key during calibration will abort the calibration, previous values will be
restored.
Summary of Contents for MDCC-4430
Page 1: ...ENABLING BRIGHT OUTCOMES MDCC 4430 NC MDCC 6530 NC User guide Coronis Fusion 4MP 6MP...
Page 5: ...5 R5911227 05 Coronis Fusion 4MP 6MP Welcome 1...
Page 8: ...R5911227 05 Coronis Fusion 4MP 6MP 8 Welcome...
Page 9: ...9 R5911227 05 Coronis Fusion 4MP 6MP Installation 2...
Page 16: ...R5911227 05 Coronis Fusion 4MP 6MP 16 Installation...
Page 17: ...17 R5911227 05 Coronis Fusion 4MP 6MP Daily operation 3...
Page 35: ...35 R5911227 05 Coronis Fusion 4MP 6MP Cleaning your display 5...
Page 37: ...37 R5911227 05 Coronis Fusion 4MP 6MP Repackaging instructions 6...
Page 41: ...41 R5911227 05 Coronis Fusion 4MP 6MP Image 6 7 Repackaging instructions...
Page 42: ...R5911227 05 Coronis Fusion 4MP 6MP 42 Repackaging instructions...
Page 43: ...43 R5911227 05 Coronis Fusion 4MP 6MP Important information 7...
Page 62: ...R5911227 05 Coronis Fusion 4MP 6MP 62 Important information...
Page 63: ......
Page 64: ...www barco com R5911227 05 2022 01 12 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...