5. Important information
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment –
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood,
concrete or ceramic tile.
If
fl
oors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 2kV for power supply
lines
± 1kV for input/ output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
< 5% U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
< 5% U
T
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
Mains power quality
should by that of a typical
commercial or hospital
environment. If the user
of the JAO15 / JAO18
requires continued
operation during power
mains interruptions, it
is recommended that
the JAO15 / JAO18
be powered from an
uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic
fi
eld
IEC 61000-4-8
3 A/m
Not applicable
2
Power frequency
magnetic
fi
elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the JAO15 /
JAO18, including cables,
than the recommended
separation distance
calculated from the
equation applicable
to the frequency
of the transmitter.
Recommended
separation distance
d = 1.2
√
P
d = 1.2
√
P
80 MHz to 800
MHz
d = 2.3
√
P
800 MHz to 2.5
Ghz
1. is the a.c. mains voltage prior to application of the test level.
2. JAO15 / JAO18 doesn’t contain susceptible components to magnetic fields
K5903250 JAO15 / JAO18 10/03/2016
23
Summary of Contents for JAO18
Page 1: ...JAO15 JAO18 User Guide Smart Terminal for the patient bedside K5903250 01 10 03 2016...
Page 4: ...Table of contents 2 K5903250 JAO15 JAO18 10 03 2016...
Page 8: ...1 Introduction 6 K5903250 JAO15 JAO18 10 03 2016...
Page 14: ...2 Installation 12 K5903250 JAO15 JAO18 10 03 2016...
Page 18: ...4 Cleaning 16 K5903250 JAO15 JAO18 10 03 2016...