5. Important information
5.4
EMC notice
When the JAO15 / JAO18 is used in combination with external peripherals, including
the JAOHandset, the test levels which are listed further below are not applicable and
the EN55022 (CISPR22) and EN55024 (CISPR24) test levels apply instead.
General information
No speci
fi
c requirement on the use of external cables or other accessories except power supply.
With the installation of the device, use only the delivered power supply or a spare part provided by the
legal manufacturer. Using another can result in a decrease of the immunity level of the device.
Electromagnetic emissions
The JAO15 / JAO18 is intended for use in the electromagnetic environment speci
fi
ed below. The customer
or the user of the JAO15 / JAO18 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment –
Guidance
RF emissions
CISPR 11
Group 1
The JAO15 / JAO18 uses RF
energy only for its internal function.
Therefore, its RF emissions are
very low and are not likely to
cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class D
Voltage
fl
uctuations/
fl
icker
emissions
IEC 61000-3-3
Complies
The JAO15 / JAO18 is suitable
for use in all establishments,
including domestic establishments
and those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
This JAO15 / JAO18 complies with appropriate medical EMC standards on emissions to, and interference
from surrounding equipment. Operation is subject to the following two conditions: (1) this device may not
cause harmful interference, and (2) this device must accept any interference received, including interfer-
ence that may cause undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following mea-
sures:
•
Reorient or relocate the receiving antenna or equipment.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The JAO15 / JAO18 is intended for use in the electromagnetic environment speci
fi
ed below. The customer
or the user of the JAO15 / JAO18 should assure that it is used in such an environment.
22
K5903250 JAO15 / JAO18 10/03/2016
Summary of Contents for JAO18
Page 1: ...JAO15 JAO18 User Guide Smart Terminal for the patient bedside K5903250 01 10 03 2016...
Page 4: ...Table of contents 2 K5903250 JAO15 JAO18 10 03 2016...
Page 8: ...1 Introduction 6 K5903250 JAO15 JAO18 10 03 2016...
Page 14: ...2 Installation 12 K5903250 JAO15 JAO18 10 03 2016...
Page 18: ...4 Cleaning 16 K5903250 JAO15 JAO18 10 03 2016...