5. Advanced operation
The available reading rooms for your Coronis Uniti are:
CR/DR/ MAMMO
Corresponds to light conditions in diagnostic
reading rooms for computed radiology, digital
radiology or mammography. This setting has the
lowest maximum ambient light.
CT/MR/NM
Corresponds to light conditions in diagnostic
reading rooms for computed tomography, magnetic
resonance or nuclear medicine scans.
Staff Of
fi
ce
Corresponds to light conditions in of
fi
ce rooms.
Clinical Viewing Room
Corresponds to light conditions in diagnostic
reading rooms for clinical viewing.
Emergency Room
Corresponds to light conditions in emergency
rooms.
Operating Room
Corresponds to light conditions in operating rooms.
This setting has the highest maximum ambient light.
To select a reading room:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > Calibration > Ambient Light
menu.
3. Enter the
Reading Room
submenu.
4. Select one of the available reading rooms and con
fi
rm.
5.20 Continuous ALC
About Continuous ALC
Continuous ALC can only be selected when the DICOM display function is selected.
Therefore, please refer to "Display functions", page 35 to correctly set the display func-
tion.
Enabling continuous ALC will continuously recalculate the DICOM display function taking the averaged
ambient light into account.
To select continuous ALC:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > Calibration > Ambient Light
menu.
3. Enter the
Continuous ALC
submenu.
4. Select
Enabled/Disabled
as desired and con
fi
rm.
5.21 Embedded QA
Overview
•
•
•
•
•
K5902079 CORONIS UNITI 21/04/2015
37
Summary of Contents for Coronis Uniti MDMC-12133
Page 1: ...Coronis Uniti User Guide MDMC 12133 K5902079 03 21 04 2015...
Page 6: ...1 Welcome 4 K5902079 CORONIS UNITI 21 04 2015...
Page 30: ...4 Daily operation 28 K5902079 CORONIS UNITI 21 04 2015...
Page 46: ...5 Advanced operation 44 K5902079 CORONIS UNITI 21 04 2015...
Page 48: ...6 Cleaning your display 46 K5902079 CORONIS UNITI 21 04 2015...