K5902079-07
Coronis Uniti
66
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment
–
guidance
Conducted RF
IEC 61000-4-6
3 Vrms (6 Vrms in ISM
bands)
150 kHz to 80 MHz
3 Vrms (6 Vrms in ISM
bands)
Portable and mobile RF
communications
equipment should be used
no closer to any part of the
Coronis Uniti, including
cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
d = 1.2
√
P
d = 1.2
√
P 80 MHz to 800
MHz
d = 2.3
√
P 800 MHz to 2.7
GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the
transmitter manufacturer
and d is the recommended
separation distance in
meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
survey,
6
should be less
than the compliance level
in each frequency range.
7
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Recommended separation distance
The Coronis Uniti is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer of the user of the Coronis Uniti can help prevent electromagnetic interference by
6:
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the Coronis Uniti is used exceeds the applicable RF compliance level above, the Coronis Uniti
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the Coronis Uniti.
7:
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Summary of Contents for Coronic Uniti
Page 1: ...ENABLING BRIGHT OUTCOMES MDMC 12133 User Guide Coronis Uniti...
Page 5: ...5 K5902079 07 Coronis Uniti Welcome 1...
Page 7: ...7 K5902079 07 Coronis Uniti Parts controls and connectors 2...
Page 12: ...K5902079 07 Coronis Uniti 12 Parts controls and connectors...
Page 27: ...27 K5902079 07 Coronis Uniti Daily operation 4...
Page 33: ...33 K5902079 07 Coronis Uniti Advanced operation 5...
Page 48: ...K5902079 07 Coronis Uniti 48 Advanced operation...
Page 49: ...49 K5902079 07 Coronis Uniti Cleaning your display 6...
Page 51: ...51 K5902079 07 Coronis Uniti Repackaging instructions 7...
Page 58: ...K5902079 07 Coronis Uniti 58 Repackaging instructions...
Page 59: ...59 K5902079 07 Coronis Uniti Important information 8...
Page 73: ...K5902079 07 Coronis Uniti 73 Important information...