7.2 Changing a measuring lamp
The Spectro LFP qb is equipped with LED modules as illuminations
placed in the dspectroal unit for reflection measurements and additional
LED modules placed in the platform under a diffusor for transmission
measurements.
These LED modules cannot be replaced by the user. They are verified
and recalibrated during the recertification process at the service centers.
To verify if the LEDs are working correctly, run the "Get Service Rport"
feature within your favorite software or the Barbieri Gateway software
and send it to the support centers.