Bante Instruments BI-650 User Manual Download Page 7

Appendix   

I   

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

0 ×  0009 - 20.00 mA correspond to conductivity value 

 

0 ×  000A- Position of decimal point for above value (20.00 mA) 

 

0 ×  000B- Measurement unit of above value (20.00 mA) 

 
 

Decimal Point Response:   

 

0 ×  0000 (Reading ×  1) 

 

0 ×  0001 (Reading ×  0.1) 

 

0 ×  0002 (Reading ×  0.01) 

 

0 ×  0003 (Reading ×  0.001) 

 
 

Measurement Unit Response:   

 

0 ×  0006: mS/cm   

 

0 ×  0007: µ S/cm 

 

0 ×  0008: ppt 

 

0 ×  0009: ppm 

     
For Example (Hexadecimal): 
PC send:

 

 

 

02  03  00  00  00  02  C4  38 

 

 

 

(Read the conductivity) 

Response: 

02  03  02  02  BC  FC  95 

 

 

 

(700) 

PC send:   

02  03  00  01  00  02  95  F8 

 

 

 

(Read the decimal point) 

Response: 

02  03  02  00  01  3D  84                           

 

 

 

(Reading ×  0.1) 

PC send:   

02  03  00  02  00  02  65  F8   

 

 

 

(Read the measurement unit) 

Response: 

02  03  02  00  07  DB  86                   

 

 

 

(µ S/cm)     

 
The result will show 70.0 µ S/cm (700 ×  0.1 µ S/cm) 
 

 

 

Electrode Maintenance 

In order to maintain an accurate measurement, the electrode needs 
cleaning and regular maintenance. 

  Remove  the conductivity  electrode  from  service and rinse  the 

 

platinum sensor on the bottom of the electrode. 

  Do not touch the platinum black coating on the sensor surface 

 

and always keep it clean.   

  If there is a build-up of solids inside the sensor, remove carefully, 

 

then recalibrate the electrode. 

  If you do not use the electrode for long periods, wipe clean with 

 

a lint-free tissue and store the electrode in a dry and cool area. 

  If your electrode is model IE-50HT, store the electrode with tap 

 

water. This sensor needs to be kept wet always. 

 

 

 

 

Appendix 

 

Troubleshooting 

 
Fault 

Cause and Corrective Action 

 

Electrode dried out.   
Soak the conductivity electrode in tap water for   
about 10 minutes. 

Measurement exceeded the maximum range.   
Check the electrode and sample. 

Drifting erratic   
readings 

Check whether electrode is contaminated, clogged 
or broken. 

 

Setting value does not match calibration solution. 
Reset the calibration value or check the calibration 
solution. 

Electrode is broken. 
Replace the conductivity electrode. 

 

Preparation of Conductivity Standard Solutions 

1.  Place the analytical grade potassium chloride (KCl) in a beaker 
 

and dry in an oven for about 3 hours at 105° C (221° F), then cool 

 

to room temperature. 

2.  Add the reagent to a 1 liter volumetric flask according to the   
 

instructions in table below. 

 

Conductivity Standard 

Reagent 

Weight 

84 µ S/cm 

KCl 

42.35 mg 

1413 µ S/cm 

KCl 

745.5 mg 

12.88 mS/cm 

KCl 

7.45 g 

111.8 mS/cm 

KCl 

74.5 g 

 

3.  Fill the distilled water to the mark, mix the solution until the   
 

reagent is completely dissolved. 

 

Calculating the Temperature Coefficient 

1.  Do not connect the temperature probe to the meter. 
2.  Press and hold the 

° C

 key to enter the temperature setting. 

3.  Press the 

 / 

 key to set the temperature to 25° C and press 

 

the 

Enter

 key to confirm. 

4.  Place the conductivity electrode into the sample solution, record 
 

the temperature value T

A

 and conductivity value C

TA

5.  Condition the sample solution and electrode to a temperature T

B

 

 

that is about 5 to 10° C different from T

A

. Record the conductivity 

 

value C

TB

6.  Calculate the temperature coefficient using the formula below. 
 

Platinum 
black coating 

 

Summary of Contents for BI-650

Page 1: ...BI 650 Industrial Conductivity Controller USER MANUAL Bante Instruments Inc...

Page 2: ...and operated only in the manner specified in this user manual Only skilled trained or authorized person should carry out installation setup and operation of meter The rear panel of meter has two scre...

Page 3: ...Display Icon Description Calibration Indicates that the meter is in the calibration mode Setup Indicates that the meter is in the setup mode ATC Indicates that the automatic temperature compensation i...

Page 4: ...cm default 0 02 S cm Analog Output High Setting range 20 00 mS cm to 0 02 S cm default 20 00 mS cm Factory Reset Reset the meter to factory default settings Note the meter must be recalibrated Enable...

Page 5: ...K 1 IE 50HT 100 S cm to 200 mS cm K 10 Conductivity Calibration TheBI 650 meter allows 1to 3pointsconductivity calibration Forbetter accuracy we recommend to perform 3 points calibration or select a s...

Page 6: ...32767 to 32767 hexadecimal PC Command Definition Length of Byte Data ID address 1 0 02 Command 1 0 03 Start address 2 0 0001 Data number 2 0 0002 CRC16 2 0 95F8 Meter Response Definition Length of Byt...

Page 7: ...Troubleshooting Fault Cause and Corrective Action Electrode dried out Soak the conductivity electrode in tap water for about 10 minutes Measurement exceeded the maximum range Check the electrode and...