Bante Instruments BI-650 User Manual Download Page 4

3   

I   

Setup 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Meter Setup 

The BI-650 meter contains an integrated setup menu for customizing 
the displayed option to meet measurement requirement. The following 
table describes the functions of each menu item. 

 
Menu Item 

Option and Description 

 

Cell Constant 

Set the cell constant to match connected conductivity 
electrode. 

 

K = 0.1, 1, 10 (default 1) 

 

Calibration Points 

Set the number of calibration points. 

 

1 to 3 points (default 1 point) 

 

Temperature Coefficient 

Set the temperature coefficient for linear temperature 
compensation. 

 

0.0 to 10.0%/° C (default 2.1) 

 

TDS Factor 

Set the default TDS conversion factor. 

 

0.1 to 1.0 (default 0.5) 

 

Temperature Unit 

 

Degrees Celsius (default) 

 

Degrees Fahrenheit 

 

Low Alarm Limit 

Setting range: 0.02 µ S/cm to 20.00 mS/cm 
(default 0.02 µ S/cm)

 

 

High Alarm Limit

 

Setting range: 20.00 mS/cm to 0.02 µ S/cm 
(default 1000 µ S/cm)

 

 

Hysteresis Value (Low)

 

Setting range: 1% to 99% (default 10%)

 

 

Hysteresis Value (High)

 

Setting range: 1% to 99% (default 1%)

 

 

Analog Output (Low)

 

Setting range: 0.02 µ S/cm to 20.00 mS/cm 
(default 0.02 µ S/cm) 

 
 

 

 

Analog Output (High) 

Setting range: 20.00 mS/cm to 0.02 µ S/cm 
(default 20.00 mS/cm)

 

 

Factory Reset 

Reset the meter to factory default settings. Note,   
the meter must be recalibrated. 

 

Enable 

 

Disable (default)   

 

 

  If the high or low alarm is enabled, the meter will be activated 

 

when the measurement exceeds specified limit. Note, this option 

 

can not enter the same values. 

  If the hysteresis is enabled, the meter will prevent rapid contact 

 

switching when the measurement is fluctuating near the set   

 

point. For example, you have set the high alarm at  20 mS/cm 

 

and hysteresis value at 1%. When the measurement overshoots 

 

the 20.2 mS/cm, the meter will activate an external device. When 

 

the measurement drops to 19.8 mS/cm, the device will switch off. 

  The default RS485 communication is 0.02 µ S/cm to 20.00 mS/cm 

 

corresponds to the 4.00 to 20.00 mA. 

 

Setting a Default Option

 

1.  In the measurement mode, press and hold the

 

 

key to enter the 

 

setup menu. 

2.  Press the 

 / 

 key to select a menu item, press the 

Enter

 key 

 

to confirm. 

 

 

 
3.  Press the 

 / 

 key to select an option or set a value, press the 

 

Enter

 key to save. 

 

 

 
 

To exit the setup menu without saving changes, press the

 

 

 

key. 

 
 

Menu   
item 
 

Page 
number 

Option 
 

Menu 
item 

 

Summary of Contents for BI-650

Page 1: ...BI 650 Industrial Conductivity Controller USER MANUAL Bante Instruments Inc...

Page 2: ...and operated only in the manner specified in this user manual Only skilled trained or authorized person should carry out installation setup and operation of meter The rear panel of meter has two scre...

Page 3: ...Display Icon Description Calibration Indicates that the meter is in the calibration mode Setup Indicates that the meter is in the setup mode ATC Indicates that the automatic temperature compensation i...

Page 4: ...cm default 0 02 S cm Analog Output High Setting range 20 00 mS cm to 0 02 S cm default 20 00 mS cm Factory Reset Reset the meter to factory default settings Note the meter must be recalibrated Enable...

Page 5: ...K 1 IE 50HT 100 S cm to 200 mS cm K 10 Conductivity Calibration TheBI 650 meter allows 1to 3pointsconductivity calibration Forbetter accuracy we recommend to perform 3 points calibration or select a s...

Page 6: ...32767 to 32767 hexadecimal PC Command Definition Length of Byte Data ID address 1 0 02 Command 1 0 03 Start address 2 0 0001 Data number 2 0 0002 CRC16 2 0 95F8 Meter Response Definition Length of Byt...

Page 7: ...Troubleshooting Fault Cause and Corrective Action Electrode dried out Soak the conductivity electrode in tap water for about 10 minutes Measurement exceeded the maximum range Check the electrode and...

Page 8: ...S cm 480 ml ECCS 1288 Conductivity standard solution 12 88 mS cm 480 ml ECCS 1118 Conductivity standard solution 111 8 mS cm 480 ml Meter Specifications Model BI 650 Conductivity Range 0 01 S cm to 2...

Page 9: ...eter is one year from the date of shipment Above warranty does not cover the electrode and standard solution Out of warranty products will be repaired on a charged basis The warranty on your meter sha...

Reviews: