Bante Instruments BI-650 User Manual Download Page 2

1   

I   

Overview 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Introduction 

Thank you for selecting the BI-650 industrial conductivity controller. 
This user manual provides a step-by-step guide to help you operate

.

 

the meter, please carefully read the following instructions before use. 
 

Environmental Conditions 

Before the installation, ensure that current environmental conditions 
meet the following requirements. 

  Relative humidity is less than 80% 

  Ambient temperature between 5° C (41° F) and 50° C (122° F) 

  No potential electromagnetic interference 

  No corrosive gas exists 

 

Packing List 

The following list describes all components of the meter. If any items 
are missing or damaged, contact the supplier immediately. 

 
BI-650 conductivity controller 
IE-50MT industrial conductivity electrode 
Conductivity standard solution 1413 µ S/cm 
 

 

 

Installation 

 

Safety Warning 

  BI-650 meter shall be installed and operated only in the manner 

 

specified in this user manual. 

  Only skilled, trained or authorized person should carry out   

 

installation, setup and operation of meter. 

 

 

  The rear panel of meter has two screw terminals for connecting 

 

the 24V DC power supply. Make sure to cut off the main power 

 

before installation and maintenance. 

  Once the power supply cable is connected to the meter, DO NOT 

 

touch any screw terminals. 

 

Installing the Meter 

 

 

 

 

 

 
 
1.1  Cut out a square hole approximately 91 ×  91 mm (3.58 ×  1.58 in.) 
 

in the mounting panel. 

1.2  Remove the mounting bracket, place the meter into the square 
 

hole. 

1.3  Replace the mounting bracket and push the meter forward until 
 

it is fully seated on the mounting plate. 

 

Installing the Electrode 

2.1  Wrap Teflon tape to the electrode body threads. 
2.2  Insert the electrode into the mounting position and slowly turn 
 

clockwise until secure. Hand tighten the electrode to prevent 

 

liquid leakage. 

 
 

Selectable mounting positions. 

 

 

 

 

   

 

 

 

 

 

   

 

 

 

 

 

   

 

 
 

The arrow represents the direction of the liquid flow. 

91 mm 

91

 mm

 

Mounting bracket 

 

 

Summary of Contents for BI-650

Page 1: ...BI 650 Industrial Conductivity Controller USER MANUAL Bante Instruments Inc...

Page 2: ...and operated only in the manner specified in this user manual Only skilled trained or authorized person should carry out installation setup and operation of meter The rear panel of meter has two scre...

Page 3: ...Display Icon Description Calibration Indicates that the meter is in the calibration mode Setup Indicates that the meter is in the setup mode ATC Indicates that the automatic temperature compensation i...

Page 4: ...cm default 0 02 S cm Analog Output High Setting range 20 00 mS cm to 0 02 S cm default 20 00 mS cm Factory Reset Reset the meter to factory default settings Note the meter must be recalibrated Enable...

Page 5: ...K 1 IE 50HT 100 S cm to 200 mS cm K 10 Conductivity Calibration TheBI 650 meter allows 1to 3pointsconductivity calibration Forbetter accuracy we recommend to perform 3 points calibration or select a s...

Page 6: ...32767 to 32767 hexadecimal PC Command Definition Length of Byte Data ID address 1 0 02 Command 1 0 03 Start address 2 0 0001 Data number 2 0 0002 CRC16 2 0 95F8 Meter Response Definition Length of Byt...

Page 7: ...Troubleshooting Fault Cause and Corrective Action Electrode dried out Soak the conductivity electrode in tap water for about 10 minutes Measurement exceeded the maximum range Check the electrode and...

Page 8: ...S cm 480 ml ECCS 1288 Conductivity standard solution 12 88 mS cm 480 ml ECCS 1118 Conductivity standard solution 111 8 mS cm 480 ml Meter Specifications Model BI 650 Conductivity Range 0 01 S cm to 2...

Page 9: ...eter is one year from the date of shipment Above warranty does not cover the electrode and standard solution Out of warranty products will be repaired on a charged basis The warranty on your meter sha...

Reviews: