Bante Instruments BI-650 User Manual Download Page 5

Calibration and Measurement   

I   

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Temperature Compensation 

The BI-650 meter is supplied with an industrial conductivity electrode. 
When the wires of electrode are connected to the meter, the display

.

 

will show ATC icon immediately. The meter is now switched to the

.

 

automatic temperature compensation mode. 
 

 

 

 

 

Temperature Calibration 

During the measurement, if the measured temperature reading differs 
from that of an accurate thermometer, the electrode needs to be   
calibrated. 
1.  Place the electrode into a solution with a known accurate   
 

temperature. 

2.  Press the 

° C

 key to enter the temperature setting. 

3.  Press the 

 / 

 key to modify the temperature value. 

4.    Press the 

Enter

 key to save. 

 

 

 

 

 

Selecting a Conductivity Electrode 

The BI-650 meter is capable of using three types of electrodes. Before 
the calibration and measurement, ensure that you have selected a   
suitable electrode according to the anticipated sample conductivity.   
The following table lists the selectable electrode and its effective   
measurement ranges. 

 
Electrode 

Measurement Range 

Cell Constant 

IE-50LT 

0.5 to 100 µ S/cm 

K = 0.1 

IE-50MT 

10 µ S/cm to 20 mS/cm 

K = 1 

IE-50HT 

100 µ S/cm to 200 mS/cm 

K = 10 

 
 
 

 

 

Conductivity Calibration 

The BI-650 meter allows 1 to 3 points conductivity calibration. For better 
accuracy, we recommend to perform 3 points calibration or select a 
standard solution closest to the sample conductivity you are measuring. 
The meter will automatically detect the standard solution and prompt 
the user to perform the calibration. The following table shows the 
default standard solution for each measurement range. 

 
Measurement Range 

Default Standard Solution 

20 to 200 µ S/cm 

84 µ S/cm 

200 to 2000 µ S/cm 

1413 µ S/cm 

2 to 20 mS/cm 

12.88 mS/cm 

20 to 200 mS/cm 

111.8 mS/cm 

 

Single Point Calibration 

Ensure that you have selected 1 point calibration in the setup menu. 
1.1  Press the 

Cal

 key, the display shows ---

 

/

 

, the meter waits 

 

for recognizing the standard solution. 

 

 

 
1.2  Rinse the conductivity electrode with distilled water, then rinse 
 

with a small amount of standard solution. 

1.3  Place the  electrode  into  the standard solution, the  meter  will 
 

automatically show the calibration standard (e.g., 1413 µ S/cm). 

 

 

 
1.4  Press the 

Enter

 key, the default calibration value begins flashing. 

1.5  Press the 

 / 

 key to set the calibration value, press the 

Enter

 

 

key to confirm and move to next digit. When the setting is   

 

completed, ensure that displayed value matches your calibration 

 

standard. 

Summary of Contents for BI-650

Page 1: ...BI 650 Industrial Conductivity Controller USER MANUAL Bante Instruments Inc...

Page 2: ...and operated only in the manner specified in this user manual Only skilled trained or authorized person should carry out installation setup and operation of meter The rear panel of meter has two scre...

Page 3: ...Display Icon Description Calibration Indicates that the meter is in the calibration mode Setup Indicates that the meter is in the setup mode ATC Indicates that the automatic temperature compensation i...

Page 4: ...cm default 0 02 S cm Analog Output High Setting range 20 00 mS cm to 0 02 S cm default 20 00 mS cm Factory Reset Reset the meter to factory default settings Note the meter must be recalibrated Enable...

Page 5: ...K 1 IE 50HT 100 S cm to 200 mS cm K 10 Conductivity Calibration TheBI 650 meter allows 1to 3pointsconductivity calibration Forbetter accuracy we recommend to perform 3 points calibration or select a s...

Page 6: ...32767 to 32767 hexadecimal PC Command Definition Length of Byte Data ID address 1 0 02 Command 1 0 03 Start address 2 0 0001 Data number 2 0 0002 CRC16 2 0 95F8 Meter Response Definition Length of Byt...

Page 7: ...Troubleshooting Fault Cause and Corrective Action Electrode dried out Soak the conductivity electrode in tap water for about 10 minutes Measurement exceeded the maximum range Check the electrode and...

Page 8: ...S cm 480 ml ECCS 1288 Conductivity standard solution 12 88 mS cm 480 ml ECCS 1118 Conductivity standard solution 111 8 mS cm 480 ml Meter Specifications Model BI 650 Conductivity Range 0 01 S cm to 2...

Page 9: ...eter is one year from the date of shipment Above warranty does not cover the electrode and standard solution Out of warranty products will be repaired on a charged basis The warranty on your meter sha...

Reviews: