background image

Specifications

Power Connection

12-pin M12/Euro-style male connector; accessory cable required

for operation

USB 2.0 Host

4-pin M8/Pico female connector; optional USB cable required for

operation of USB flash drive

Ethernet Connection

4-pin M8/Pico male connector

Remote Display Connection

8-pin M12/Euro-style female connector; accessory cable required

for remote display

Power Requirements

Voltage: 10 V dc to 30 V dc

Current: 1 A maximum (exclusive of I/O load)

Output Configuration

NPN or PNP, software selectable

Demo Mode

Full tool functionality on canned images

Sensor Lock

Optional password protection

Integrated Ring Light

Models with red, blue, green, infrared, white, ultraviolet or no

integrated ring light

Output Rating

150 mA

External Strobe Output

+ 5 V dc

Acquisition

60 fps (frames per second) maximum, with full FOV

Exposure Time

0.1 ms to 1.049 s

Imager

1/3 inch CMOS 752 × 480 pixels; color or grayscale, depending on

model

Adjustable Field of View (FOV)

Lens Mount

Micro Video Lens models: M12 × 1 mm thread; micro video lens 4.3

mm, 6 mm, 8 mm, 12 mm, 16 mm, 25 mm

C-Mount models: Standard C-mount (1 inch-32 UN)

Construction

Black PBT sensor housing; die cast zinc back cover; acrylic window

Integrated Display Weight: 0.36 kg (0.80 lbs)

Remote Display Weight: 0.41 kg (0.90 lbs)

Environmental Rating

IEC IP67, micro video lens models only

Operating Conditions

Integrated Display Stable Ambient Temperature: 0 °C to +45 °C

(+32 °F to +113 °F)

Remote Display Stable Ambient Temperature: 0 °C to +40 °C

(+32 °F to +104 °F)

Certifications

Banner Engineering Corp. Limited Warranty

Banner Engineering Corp. warrants its products to be free from defects in material and workmanship for one year following the date of shipment. Banner Engineering Corp. will repair or

replace, free of charge, any product of its manufacture which, at the time it is returned to the factory, is found to have been defective during the warranty period. This warranty does not

cover damage or liability for misuse, abuse, or the improper application or installation of the Banner product.
THIS LIMITED WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED (INCLUDING, WITHOUT LIMITATION, ANY WARRANTY OF

MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE), AND WHETHER ARISING UNDER COURSE OF PERFORMANCE, COURSE OF DEALING OR TRADE USAGE.
This Warranty is exclusive and limited to repair or, at the discretion of Banner Engineering Corp., replacement. IN NO EVENT SHALL BANNER ENGINEERING CORP. BE LIABLE TO

BUYER OR ANY OTHER PERSON OR ENTITY FOR ANY EXTRA COSTS, EXPENSES, LOSSES, LOSS OF PROFITS, OR ANY INCIDENTAL, CONSEQUENTIAL OR SPECIAL DAMAGES

RESULTING FROM ANY PRODUCT DEFECT OR FROM THE USE OR INABILITY TO USE THE PRODUCT, WHETHER ARISING IN CONTRACT OR WARRANTY, STATUTE, TORT,

STRICT LIABILITY, NEGLIGENCE, OR OTHERWISE.
Banner Engineering Corp. reserves the right to change, modify or improve the design of the product without assuming any obligations or liabilities relating to any product previously

manufactured by Banner Engineering Corp. Any misuse, abuse, or improper application or installation of this product or use of the product for personal protection applications when the

product is identified as not intended for such purposes will void the product warranty. Any modifications to this product without prior express approval by Banner Engineering Corp will

void the product warranties. All specifications published in this document are subject to change; Banner reserves the right to modify product specifications or update documentation at

any time. Specifications and product information in English supersede that which is provided in any other language. For the most recent version of any documentation, refer to: 

www.bannerengineering.com

.

 

 

iVu Plus TG and Color Gen2 Image Sensors

 

©

 Banner Engineering Corp. All rights reserved

 

Summary of Contents for Color Gen2

Page 1: ...N Electrostatic Discharge Avoid the damage that electrostatic discharge ESD can cause to the Sensor Always use a proven method for preventing electrostatic discharge when installing a lens or attachin...

Page 2: ...re Color Description Direction 1 White Output 1 Output 2 Brown 10 V dc to 30 V dc Input 3 Green Output 2 Output 4 Yellow Strobe Out 5 V dc only Output 5 Gray Remote Teach Input 6 Pink External Trigger...

Page 3: ...sensor inspection The green annotations indicate the object passes and the red annotations indicate a failure This sample inspection failed as shown by the red X next to the Display mode button Regio...

Page 4: ...Count Sort Saved Patterns Percent Match Rotation Range Pass Criteria Inspection Name Set Name ID 3 Not available in Match and Sort Inspection Timeout Select Timeout Average Color 1 ROI and Mask Inten...

Page 5: ...cons Icon Description The Main Menu icon is on the bottom left corner of the display on the Home screen Click this icon to access sub menus that are used to set up the iVu The Inspection menu icon is...

Page 6: ...hrough and select a circular elliptical or rectangular shaped mask The Elliptical Mask icon is on the left side of the display when a mask is selected Click this icon to cycle through and select a cir...

Page 7: ...actual objects a Go to Imager White Balance b Move and adjust the white balance Region of Interest ROI to surround a white or gray object in the Field of View FOV If there is none place a piece of wh...

Page 8: ...e separately Adjust the Focus on a C Mount Lens Model 1 Remove the lens enclosure 2 Adjust the focus while monitoring the focus number To ensure the best image adjust the focus until the focus number...

Page 9: ...Sensor Use a Sort sensor to recognize and sort up to ten different patterns within the same inspection Teach each reference pattern during setup Store reference patterns in one of ten pattern memory l...

Page 10: ...milliseconds Note If the interval is less than the inspection time missed triggers will occur Free Run the iVu automatically runs continuous inspections Industrial Ethernet Only only Trigger commands...

Page 11: ...l Ethernet A bi directional communication channel that allows the user to control the device and access device results using Ethernet IP 1 Modbus TCP PCCC or PROFINET 2 protocol Data Export Used to ex...

Page 12: ...it is returned to the factory is found to have been defective during the warranty period This warranty does not cover damage or liability for misuse abuse or the improper application or installation...

Reviews: