Manual Grinding Wheel 2928 00
9
Problems and solutions
In the following chapter possible causes for faults and work needed for their
resolution are described.
For problems and faults that cannot be resolved using the following instructions,
contact your authorised dealer or the manufacturer. For Germany see the service
address on page 2.
9.1 Troubleshooting guide
Personnel: Specialized personnel
Fault description
Cause
Remedy
Motor does not
turn
Instrument
switched off
Switch on the device (Chapter 7.2.4 ‘Swit-
ching the instrument on/off’ on page 22).
Instrument not
connected to the
power supply
Connect to the power supply (Chapter 6.5
‘Connect the unit to the power supply’ on
page 19).
Power failure
Check the fuse of the power supply of the
healthcare establishment.
Defective instru-
ment fuse
Replace the instrument fuse on the back of
the instrument in the instrument inlet recep-
tacle fuse holder.
Instrument inlet
connector loose
Check instrument inlet connector is plugged
firmly in.
Grinding speed is
inadequate
Grinding wheel
blocked up with
grinding material
Remove material from grinding wheel with the
cleaning stone (Chapter 8.4 ‘Cleaning the
grinding wheel (removal)’ on page 26).
Grinding wheel is
dry although tap
water is dripping
from the water
valve
Sponge is
blocked up with
grinding sludge
Clean the sponge or replace if necessary
(Chapter 8.3 ‘Cleaning sponges’ on page
24).
Water is running
down the sides of
the upper outer
housing if tap
water is dripping
from the water
valve
Sponge is
blocked up with
grinding sludge
Clean the sponge or replace if necessary
(Chapter 8.3 ‘Cleaning sponges’ on page
24).
Water leaks from
the bottom of the
manual grinding
wheel
Collection contai-
ner is full
Empty the collection container (Chapter 7.2.1
‘Check water collection container’ on page
20).
Overflow hose
on the collec-
tion container is
kinked or blocked
Check the position of the overflow hose on
the operator‘s side or remove the blockage.
Problems and solutions