background image

Instructions for Use

!

1

Ultrasound since 1955

23346 GB/2015-03

3 4

Ultrasonic and rinsing baths for aqueous 

fl

 uids

Valid for: 

RM 110 UH, RM 110 U, RM 110 H, RM 110

RM 180 UH, RM 180 U, RM 180 H, RM 180

RM 210 UH, RM 210 U, RM 210 H, RM 210

Summary of Contents for SONOREX TECHNIK RM 110 UH

Page 1: ...or Use 1 Ultrasound since 1955 23346 GB 2015 03 3 4 Ultrasonic and rinsing baths for aqueous fluids Valid for RM 110 UH RM 110 U RM 110 H RM 110 RM 180 UH RM 180 U RM 180 H RM 180 RM 210 UH RM 210 U R...

Page 2: ...en the translation and the original of this document the original version will take prec edence BANDELIN accepts no responsibility or liability for damages caused by improper handling or usage contrar...

Page 3: ...nauthorised alterations modifications both the warranty claim and the conformity will no longer be valid If service is required please contact the specialist dealer in charge or the manufacturer Symbo...

Page 4: ...nings and safety precautions 12 2 Preparation 13 2 1 Scope of delivery 13 2 2 Set up assembly 13 2 3 Start up 14 3 Operation 15 3 1 Operating elements 15 3 1 1 Ultrasound 15 3 1 2 Heating 16 3 2 Misce...

Page 5: ...service 24 6 4 2 Decontamination certificate 25 6 4 3 Replacing fuses 25 7 Accessories 26 7 1 Required accessories 26 7 2 Optional accessories 27 7 3 Chemical preparations Recommendations 29 8 Consum...

Page 6: ...z Easy to clean stainless steel housing 2 Filling level mark for safe filling 3 Sprinkle tube 4 and weir overflow 5 with ball valve to produce a liquid circuit Height adjustable legs 6 for a secure an...

Page 7: ...gas content 1 2 Purpose SONOREX TECHNIK ultrasonic baths are intended for the sonication of aqueous liquids They work on the basis of low frequency ultrasound and can be used in versatile ways Their...

Page 8: ...side of the tank Serial number see type plate on rear of device Degree of protection IP 32 according to DIN EN 60529 Environmental conditions pursuant to EN 61 010 1 Overvoltage category II Degree of...

Page 9: ...outlet G inside G 1 outside Heating power W 4800 4800 Fuses heating T12A T12A HF frequency kHz 25 or 40 25 or 40 Fuses generator T6A T6A Peak ultrasonic output W 4000 4000 HF power Weff 1000 1000 Wei...

Page 10: ...G inside G 1 outside Heating power W 7200 7200 Fuses heating T12A T12A HF frequency kHz 25 or 40 25 or 40 Fuses generator T6A T6A Peak ultrasonic output W 2 4000 2 4000 HF power Weff 2 1000 2 1000 We...

Page 11: ...G inside G 1 outside Heating power W 7200 7200 Fuses heating T12A T12A HF frequency kHz 25 or 40 25 or 40 Fuses generator T6A T6A Peak ultrasonic output W 2 4000 2 4000 HF power Weff 2 1000 2 1000 We...

Page 12: ...assed on to others the operating instructions with the safety instructions must also be handed over Operation and transport Observe ambient and set up conditions see chapter 1 4 Only plug in the ultra...

Page 13: ...ssembly Place the ultrasonic rinsing bath on a firm level and dry surface In doing so observe the maximum weight of the tank including liquids For net weight see technical data chapters 1 4 1 to 1 4 3...

Page 14: ...e limit stop or back it off by rotating it slightly to the left this will cause the PTFE tape to lose its sealing effect 5 Close ball valve turn the lever 5 upwards 2 3 Start up Thoroughly rinse the u...

Page 15: ...rasonic bath does not turn off automatically to switch it off turn the knob to the right back to 0 Notes Even while turned off the ultrasonic bath may remain connected to the mains It can be disconnec...

Page 16: ...The yellow control light goes out when the set temperature is reached To turn it off turn the knob left back to C Notes The heating system works independently from the ultrasound The heating automati...

Page 17: ...ce may switch off as a result of evaporation for example In such cases the ultrasound and possibly the heating must be switched off controller on 0 Once the tank is filled up the ultrasound and the he...

Page 18: ...ing tank may not exceed a maximum operating temperature of 100 C Cover the ultrasonic bath when in continuous operation so that not too much bath liquid evaporates For safety reasons ultrasonic cleani...

Page 19: ...Check that the objects to be treated are fully covered with fluid With every object inserted the fill level is to be controlled For cleaning tasks Place the objects to be cleaned in the appropriate a...

Page 20: ...onication results Before the next sonication verify the service life see chapter 4 3 2 of the bath fluid Heed the specifications of the preparation manufacturer If necessary empty the oscillating tank...

Page 21: ...by DR H STAMM GmbH are prepared pursuant to the regulations of the German Washing and Cleansing Agents Act are biodegradable and may be discharged into the sewerage as working solutions Strongly acidi...

Page 22: ...Do not use steel wool scrapers or graters for cleaning maintenance The free movement and function of the float used for fill level control on the front left side must be checked from time to time Met...

Page 23: ...e ultrasound effect For this check it is recommended that a foil test be conducted semi annually Ordinary aluminium foil is used to conduct the test Next a comparison is made with previously generated...

Page 24: ...c bath can be readily operated without heating Slight erosion visible on the bottom of the bath Natural wear Ultrasonic bath OK Strong erosion marks appear at the tank bottom lead to loss of power Any...

Page 25: ...may only be carried out by authorised qualified personnel or by the manufacturer The manufacturer assumes no liability for unauthorised work performed on the ultrasonic bath The mains plug must be pul...

Page 26: ...1 Required accessories The baskets and basket holder for example are required accessories Do not stand or lay any objects directly on the tank bottom Series RM 110 Series RM 180 Series RM 210 Stainle...

Page 27: ...iltration with prefilter and main filter including connection set FA 110 8611 FA 180 8612 FA 210 8613 Oil separator incl connection set OX 110 8603 OX 180 8604 OX 210 8605 DI water treatment device in...

Page 28: ...very oily or greasy parts Impurities that rise to the surface of the tank are led via the weir into the oil separator where they are separated by gravity DI water treatment device WA To be connected...

Page 29: ...e best cleaning results can be achieved when using the following TICKOPUR preparations Contamination Objects to be cleaned Cleaning concentrate Litres Order No Universal cleaner General contamina tion...

Page 30: ...ease contaminants etc Metals including burnished metals glass ceramics plastics rubber etc Especially for galvanic laser and analytical appli cations Dilute with DI water TICKOPUR R 32 Special cleaner...

Page 31: ...l ishing and lapping residues etc Steel stainless steel glass ceramics plastics rubber Not for light metals tin zinc TICKOPUR R 60 Intensive cleaner phosphate free strongly alkaline pH 12 8 1 applicat...

Page 32: ...ot applicable 9 Taking the unit out of service If the ultrasound rinsing bath no longer works it must be appropriately disposed of Some electrical components are considered to be toxic waste 10 Key wo...

Page 33: ...e bath during the cleaning process improving the cleaning effect The modular system allows for individual assembly of these items depending on the cleaning task A significant advantage is that the sys...

Page 34: ...ies must be cleaned pursuant to current laws and regulations and if necessary must also be disinfected with a surface disinfection agent listed by the VAH Device type Serial number Accessories Device...

Page 35: ...sories found in this package have been cleaned and or disinfected pursuant to current laws and regulations and that the information provided in this declaration is correct and complete Company Institu...

Page 36: ...Note The Instructions for Use in this and other languages as well as further information can be found on the enclosed CD...