Mount the separately-packaged ball valves for each tank as follows:
1. Screw the counter nut onto the threaded sleeves 1 at the
tank outlet (3-way ball valve) and tank over
fl
ow (ball valve),
then seal with the white PTFE tape provided, applying
several turns of tape to the right.
2. Tape the hose socket threads 2 with PTFE sealing tape.
3. Screw the hose sockets 2 into the ball valve outlets 3, use
the appropriate spanner.
4. Screw the ball valve 3 onto the threaded sleeves and
tighten with the counter nut 4.
Caution:
Do not screw the ball valve up to the limit stop
or back it off by rotating it slightly to the left, this will cause
the PTFE tape to lose its sealing effect.
5. Close ball valve, turn the lever 5 upwards.
2.3 Start-up
Thoroughly rinse the ultrasonic or rinsing bath's oscillating tank with water before its
fi
rst use.
Verify that the control buttons are in the "off" position, i.e. the switch indicator is at
the top, then connect the bath to the mains.
Conduct function test on the ultrasonic bath – brie
fl
y plug in the ultrasound
(maximum of 1 to 2 seconds), a hissing noise should be heard. Set to "0" once
again.
Hang accessories in the ultrasonic bath and place lid on top.
Connect peripheral devices if required – see additional documentation.