AP0318b
AP0315a
ELECTRICAL CONNECTION (Immersion probe CP2 or CP5)
10 to 30 V DC
100 to 240 V AC
Shield (supply –)
L(+) N(+)
Fuse
16
14
Green (CP5 only)
Service wire
White (signal)
Brown ()
may not be
connected
8
2
7
6
1
3
4
Power Supply
OC V+
UN RX
V+
TURBISWITSCH GS5
Immersion
probe CP...
Limit value >>
Limit value <<
Fault
9
11
12
13
15
Signalling and Control Unit
SWITCHING LOGIC
Switching Logic
Limit value
exceeded
and
No
fault
Below the
limit value
and
No
fault
Measurement
impossible
and
fault
9
11
12
13
15
9
11
12
13
15
9
11
12
13
15
Switching logic with increased reliability
The contacts open when one of the following events occur:
Contact 9-11 opens when the limit value is exceeded.
Contact 9-12 opens when measurements are below the limit value.
Contact 13-15 opens in the event of a fault.
Please note:
A connected PLC must react to the opening contact 13-15 in order to ensure a wire break monitoring.
If a fault occurs or in the event of power failure, all the contacts (9,11,12,13,15) are open.
TUR
19-08-2021
M-410.03-EN-AA
410-03
/6
Turbidity Limit Value
Evaluation Unit
TURBISWITCH GS5