TURBISWITCH GS5

Turbidity Limit Value Evaluation Unit

Evaluation Unit

Measuring cells (in-line installation)

Immersion Probes

SAFETY PRECAUTIONS

• Installation, initial start-up and maintenance may only be performed by trained personnel!

• The device may only be connected to supply power which complies with the specifications in the technical data and on the serial plate!
• The device must be disconnected from all sources of power during installation and maintenance work!
• The device may only be operated under the conditions specified in the operating instructions!

DESCRIPTION

The TURBISWITCH GS5 evaluation unit in conjunction with the associated transducers is an optical turbidity measurement for determining
the solid content in a liquid medium.
The turbidity signal is read out as an adjustable limit value via the evaluation unit.

The turbidity measurement is based on the optical absorption method, i.e. it reacts to light loss due to undissolved solids in the medium. Due to
clocked infrared light, the measurement is insensitive to external light.
If the measured turbidity value is exceeding or below the trigger point value, the output relay in the TURBISWITCH GS5 is actuated in

accordance with the settings.

Attention: Operation of the transducers of the TRUBOMAT series (TT-GS, TR-GS and CP1) at the evaluation unit
TURBISWITCH GS5 is not possible!

TECHNICAL DATA

Supply power

100...240V AC / 50...60Hz (TURBISWITCH  GS5 G) or
10...30V DC and 12...24V AC (TURBISWITCH GS5 D)

Power consumption

1...5W

Ambient temperature

-10...+45°C

Output relays

2x potential free limit value contact (normally open)
(turbidity value is exceeding or below the trigger point value)

1 x potential free fault contact (normally open)

All contacts are open when supply power is switched off!

Output relay switching capacity

250V AC, 3A / 30V DC, 1A

Note: Contacts are not protected against overload! Use external protective device!

Housing dimensions

22.5x100x122mm

Top-hat rail dimensions

35x7.5mm (DIN EN 60715)

Protection

IP40

Connector terminals

Screw terminals, max. 1.5mm²

Limit value

0…100% within 3 measuring ranges (depending on the solid content)
LOW (5% steps); MEDIUM (2% steps); HIGH (1% steps)

Reset hysteresis

Adjustable 1…25% of the selected limit value

Cable length

Max. 100m between sensor and evaluation unit

Display

2½-place LED, 5x7 pixel matrix display

Settings

Rotary/pushbutton selector on front panel

Switching delay

Adjustable 0.1…9.9 seconds

CE Mark: The device fulfills the legal requirements of applicable EU-guidelines.

TUR

19-08-2021

M-410.03-EN-AA

410-03

/1

Turbidity Limit Value

Evaluation Unit

TURBISWITCH GS5

22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL

Tel

Fax

+33 (0)1 30 25 83 20

+33 (0)1 34 10 16 05

Web

E-mail

www.bamo.eu

[email protected]

INTERNATIONAL

Summary of Contents for TURBISWITCH GS5

Page 1: ...OMAT series TT GS TR GS and CP1 at the evaluation unit TURBISWITCH GS5 is not possible TECHNICAL DATA Supply power 100 240V AC 50 60Hz TURBISWITCH GS5 G or 10 30V DC and 12 24V AC TURBISWITCH GS5 D Po...

Page 2: ...shed by simply turning the rotary pushbutton selector The selected hysteresis value is automatically skipped in this case 100 in the LOW range is followed by 0 in the MEDIUM range 100 in the MEDIUM ra...

Page 3: ...iefly displayed The display is then limit value display INSTALLING A MEASURING CELL IN A PIPING SYSTEM IN LINE Installation of in line measuring cell The Measuring cells TURBISWITCH GA1 GA5 GA11 can b...

Page 4: ...TURBISWITCH CP2 and CP5 transducers are designed as hanging probes or for use with an extension tube It can be installed in slurry tanks basins or open wells The probe should be installed so that it c...

Page 5: ...eiver TT HDR and TR HDR or TT HDR5 and TR HDR5 Emitter Receiver IR Light TT HDR Emitter TURBISWITCH GS5 TR HDR Receiver 10 to 30 V DC 100 to 240 V AC L N Fuse 16 14 8 2 6 1 3 4 OC V RX V Limit value L...

Page 6: ...13 15 Switching logic with increased reliability The contacts open when one of the following events occur Contact 9 11 opens when the limit value is exceeded Contact 9 12 opens when measurements are b...

Page 7: ...NG THE THRESHOLD LIMIT VALUE AND COMMISSIONING Adjustment takes place after installation and electrical connection If problems arise see Troubleshooting All steps must be carried out in chronological...

Page 8: ...is so high that no light is reaching the receiver HDR5 emitter blinks red Defective emitter Replace the emitter TR appears on the display The receiver has malfunctioned or it is not connected Check t...

Page 9: ...mescale remover The side of the glass disc marked with a dot must be installed that it faces the medium Check the seals O rings for damage and replace them if necessary before start up The frequency o...

Reviews: