Bamo TURBISWITCH GS5 Manual Download Page 4

DP0006a

INSTALLING A MEASURING CELL IN A PIPING SYSTEM (IN-LINE) / (Continuation)

To assure a reliable measurement please observe the following:

• The optical windows must be clean!
• The formation of gas bubbles in the medium should be prevented to avoid false measurements!
• The emitter and the receiver must always be aligned with the horizontal axis of the pipe when installed horizontally in order to avoid

incorrect measurements due to deposits on the glass discs (see picture below)!

E

R

R

E

• Only a completely filled pipeline guarantees a correct measurement!

• It may be advisable to provide for back pressure by means of restriction, or to install a trap.
• Dried-on residues on the glass discs distort measurements, e.g. after an operating process was in standby.

• It is recommended to include straight pipe distances, upstream and downstream the measuring cell, of at least 3 to 5 times the diameter of

the pipe.

• When mounted in a trap, stop valves or/and drain valve should be installed for easy maintenance and removing the measuring cell.
• Use the supplied mounting kit (key tool and suction cup) in order to remove and install the glass discs. Observe torque! (Torque spanner

Turbiclick 5.7 optionally available, please contact us).

INSTALLING AND CONNECTING THE IMMERSION PROBE (CP2 or CP5)

Installing the immersion probe

The TURBISWITCH CP2 and CP5 transducers are designed as hanging probes or for use with an extension tube.
It can be installed in slurry tanks, basins or open wells.

The probe should be installed so that it can be easily removed for cleaning.
Gas bubbles cause false measurements!
The frequency of cleaning routines depends on operating conditions.
Do not scratch the optics when cleaning the sensors!

The following measuring ranges apply to the TURBISWITCH CP2 and CP5 transducers:

Minimum detectable turbidity in the LOW range: starts at roughly 100...300FAU
Maximum detectable turbidity in the HIGH range: approx. 30000FAU (corresponds to approx. 100g/l SiO₂)
Maximum detectable turbidity in the MEDIUM range: 10% of the HIGH range

Maximum detectable turbidity in the LOW range: 1% of the HIGH range

Installation of the immersion probe TURBISWITCH CP2 or CP5 variant ZK or Z0

CP2 PVC

CP5 PP

The Z0 version is supplied without fitting accessories.
The ZK version is supplied with an angle bracket and its cable gland.
Please note: Caution with powerful flow, turbulences the cable must not be

overstretched!
In such cases, a cable or a rod may be used to secure the sensor (fixed on to the
two holes in the base of the sensor).
M6 threads can be tapped into the holes if necessary.

TUR

19-08-2021

M-410.03-EN-AA

410-03

/4

Turbidity Limit Value

Evaluation Unit

TURBISWITCH GS5

22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL

Tel

Fax

+33 (0)1 30 25 83 20

+33 (0)1 34 10 16 05

Web

E-mail

www.bamo.eu

[email protected]

INTERNATIONAL

Summary of Contents for TURBISWITCH GS5

Page 1: ...OMAT series TT GS TR GS and CP1 at the evaluation unit TURBISWITCH GS5 is not possible TECHNICAL DATA Supply power 100 240V AC 50 60Hz TURBISWITCH GS5 G or 10 30V DC and 12 24V AC TURBISWITCH GS5 D Po...

Page 2: ...shed by simply turning the rotary pushbutton selector The selected hysteresis value is automatically skipped in this case 100 in the LOW range is followed by 0 in the MEDIUM range 100 in the MEDIUM ra...

Page 3: ...iefly displayed The display is then limit value display INSTALLING A MEASURING CELL IN A PIPING SYSTEM IN LINE Installation of in line measuring cell The Measuring cells TURBISWITCH GA1 GA5 GA11 can b...

Page 4: ...TURBISWITCH CP2 and CP5 transducers are designed as hanging probes or for use with an extension tube It can be installed in slurry tanks basins or open wells The probe should be installed so that it c...

Page 5: ...eiver TT HDR and TR HDR or TT HDR5 and TR HDR5 Emitter Receiver IR Light TT HDR Emitter TURBISWITCH GS5 TR HDR Receiver 10 to 30 V DC 100 to 240 V AC L N Fuse 16 14 8 2 6 1 3 4 OC V RX V Limit value L...

Page 6: ...13 15 Switching logic with increased reliability The contacts open when one of the following events occur Contact 9 11 opens when the limit value is exceeded Contact 9 12 opens when measurements are b...

Page 7: ...NG THE THRESHOLD LIMIT VALUE AND COMMISSIONING Adjustment takes place after installation and electrical connection If problems arise see Troubleshooting All steps must be carried out in chronological...

Page 8: ...is so high that no light is reaching the receiver HDR5 emitter blinks red Defective emitter Replace the emitter TR appears on the display The receiver has malfunctioned or it is not connected Check t...

Page 9: ...mescale remover The side of the glass disc marked with a dot must be installed that it faces the medium Check the seals O rings for damage and replace them if necessary before start up The frequency o...

Reviews: