9.6

mA OUTPUT SETTINGS FOR THE MEASUREMENT

Go to the menu OUTPUT mA FLOW

HIGHER 0360 m3

Enter the measured value for which the output is 20 mA, then confirm.

LOWER 0000 m3

Enter the measured value for which the output is 0 or 4 mA, then confirm.

OUTPUT 4/20 mA or 0/20 mA

Select the output signal type, then confirm.

SAVE ?

To record the settings, press the icon SAVE

9.7

mA OUTPUT SETTINGS FOR THE TEMPERATURE

Go to the menu OUTPUT mA TEMP.

160.0 °C

Enter the temperature value for which the output is 20 mA, then confirm.

LOWER +000.0 °C

Enter the temperature value for which the output is 0 or 4 mA, then confirm.

OUTPUT 4/20 mA or 0/20 mA

Select the output signal type, then confirm.

SAVE ?

To record the settings, press the icon SAVE

9.8

SIMULATION ON RELAYS

With the menu FORCED RELAY, it is possible to manually test (force) relays S1, S2, S3 and S4.

Last one, S4, is N.O. by default.
The test begins by relay S1.
The respective relay can be switched from OFF (open) to ON (closed).
Confirm to go to the next relay, then to go back to previous menu.

9.9

RESET THE DAILY VOLUME TO ZERO

Go to the menu RAZ Volume Jour = Rest daily volume to zero.

RESET D.V. YES / NO

Sélect YES to reset the counter to 0 on daily volume, then confirm, and SAVE.

DEB

21-12-2018

M-759.03-EN-AB

759-03

/11

Flow calculator

BAMOPHAR 759

22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL

Tel

Fax

+33 (0)1 30 25 83 20

+33 (0)1 34 10 16 05

Web

E-mail

www.bamo.eu

[email protected]

INTERNATIONAL

Summary of Contents for BAMOPHAR 759

Page 1: ...MANUAL DEB 21 12 2018 M 759 03 EN AB 759 03 1 Flow calculator BAMOPHAR 759 22 Rue de la Voie des Bans Z I de la gare 95100 ARGENTEUIL Tel Fax 33 0 1 30 25 83 20 33 0 1 34 10 16 05 Web E mail www bamo...

Page 2: ...UREMENT PARAMETERS 9 2 10 THRESHOLD S1 AS ON OFF MODE 9 3 10 THRESHOLD S3 SET AS ON OFF MODE AND ASSIGNMENT TO EXTERNAL SENSOR 9 4 10 SETTING THE SAMPLER CONTROL Relay S2 9 5 10 SETTING THE TEMPERATUR...

Page 3: ...ay S3 Adjustable hysteresis from 0 to 100 Adjustable delay from 0 to 9999 s Sampler control S2 Programming in m on volume counter 1 contact N O potential free Adjustable timer from 0 to 9999 s Pulse o...

Page 4: ...ccess to slot by removing the cover Signal input 4 20 mA Signal input 4 20 mA DEB 21 12 2018 M 759 03 EN AB 759 03 4 Flow calculator BAMOPHAR 759 22 Rue de la Voie des Bans Z I de la gare 95100 ARGENT...

Page 5: ...l 19 S2 20 Threshold S1 N O contact Measurement of flow rate or temperature 21 S1 22 Threshold S3 N O contact Flow rate or temperature measurement 23 S3 Available for assignment to external sensor bec...

Page 6: ...ly Brown Brown Blue Black Brown Blue Black Output Inductive sensor 194 831 NPN N C Connection of a N C flow controller NPN code 194831 DEB 21 12 2018 M 759 03 EN AB 759 03 6 Flow calculator BAMOPHAR 7...

Page 7: ...symbol appears if relay is set up Designation of each measuring systems Measured values display Main unit flow and temperature Extension example with pH and temperature Monitor A is shown in yellow ma...

Page 8: ...ed CONSULTATION MODE Query mode RETURN KEY Return to previous screen CONFIRM Access to the next line of menu SAVE Save settings 7 DISPLAY SETTINGS Note To access to the display settings BAMOPHAR must...

Page 9: ...n we find the serial number The serial number SN is written on the identification label of the BAMOPHAR It appears as well in MENU icon i INFORMATION 9 SETTINGS 9 1 SETTING OF MEASUREMENT PARAMETERS C...

Page 10: ...e previous chapter THRESHOLD S3 has a second function assignment to an external sensor see further on ALARM 3 ON OFF Select the ON mode in order to activate the relay Then confirm EXTERN NO YES NO Set...

Page 11: ...0 mA or 0 20 mA Select the output signal type then confirm SAVE To record the settings press the icon SAVE 9 8 SIMULATION ON RELAYS With the menu FORCED RELAY it is possible to manually test force rel...

Reviews: