 III - PROPER CABINET USE 

 

CAUTION 

A Biological Safety Cabinet is a valuable supplement to, but not a replacement for, good laboratory 
technique and safe practice. 

If the operator does not operate the cabinet correctly, it may not provide an adequate protective 
barrier. To ensure personnel protection, product protection, and environmental protection the 
cabinet must be operated per the manufacturer’s instructions. 

The facility industrial hygienist or safety officer shall ensure that: 

 

The cabinet is appropriate for all operations and procedures to be performed. 

 

All operators are thoroughly trained and competent regarding cabinet operation and all procedures they 
are required to perform. 

 

That the cabinet operation, procedures, and operators are monitored at regular intervals to ensure that 
safety is maintained. 

A good reference source is The Biosafety In Microbiological and Biomedical Laboratories (BMBL) 5

th

 edition 

published by the U.S. Department of Health & Human Services as HHS Publication No. (CDC) 21-1112 as an 
advisory document for safe work practices. 
 
Baker biosafety cabinets are designed for continuous operation. It is recommended that the blower be left on at 
all times to provide containment and keep the interior work area clean and free of particulates.  
 

Operation and Controls 

The operator controls with indicators are arranged on the front electrical panel of the cabinet. A number of 
switches are arranged in a single membrane switch assembly. [Reference Figure 3] 

 
 
 
 
 
 
 
 
 

 

 

          

Figure 3: Operator Controls 

Blower 

This pushbutton switch controls the cabinet blower. Pressing the Blower pushbutton switch will turn the blower 
ON. A green indicator light located below the switch will illuminate when the switch is on. Pressing the Blower 
pushbutton switch when the blower is running will turn the blower OFF. If the cabinet blower is turned off, the 
Fluorescent Lights will turn off. 

ALARM

Alarm Mute / Reset 

 Sash Level Alarm 

Blower 

On / Off  

Auxiliary Outlets 

On / Off 

Fluorescent Light 

On / Off 

UV Light 

On / Off 

Indicator 

Lights 

Summary of Contents for SterilGARD e3 Series

Page 1: ...B i OPERATOR S MANUAL Class II Type A2 Biological Safety Cabinet MODELS SG404 SG504 SG604 SG404 INT SG504 INT SG604 INT SG404 J SG504 J SG604 J Revision Date Description B 11 12 14 Changed Warranty St...

Page 2: ...ring a contamination free atmosphere In addition to the high quality you expect from all Baker equipment this model has been designed to provide the lab user with the most comfortable work environment...

Page 3: ...ea Lighting 4 Electronic Ballast 4 Ground Fault Circuit Interrupter 115V only 4 Laboratory services 4 Alarms 4 Timers 4 Padded Bypass Armrest 5 UniPressure Plenum 5 Motor Blower Assembly 5 Cable Port...

Page 4: ...increment programming 10 1 hour increment programming 10 Cable Ports 10 Cable Ports continued 11 Start up Procedure 11 Working in the Cabinet 12 Using Ancillary Equipment 13 Reacting to Spills 13 Clea...

Page 5: ...supply air passes through a diffuser but some of the air is directed down the back side of the view screen creating a high velocity air curtain at the front access opening The HEPA filtered supply air...

Page 6: ...ILTER POSITIVE PRESSURE PLENUM SUPPLY HEPA FILTER REMOVABLE WORK SURFACE AND SUPPORTS FLUORESCENT LAMP LIGHT CANOPY SLIDING VIEWSCREEN 10 DEGREE SLOPE SUPPLY DIFFUSER FILTER PROTECTOR AUXILIARY OUTLET...

Page 7: ...cabinet has a manual internal balancing damper that compensates for down flow and exhaust intake imbalances WARNING This damper must only be adjusted by a qualified technician who has had proper train...

Page 8: ...The SterilGARD e3 standard view screen access opening is 8 203 mm Other optional factory configurations available with NSF listing are 10 254 mm and 12 305 mm access openings Work Area Lighting The w...

Page 9: ...the recorded value in the most recent certification report However if drastic change is noted this should be cause for investigating the reason This device is not intended to be used for air flow set...

Page 10: ...rnia OSHPD pre approval FlexAIR Canopy Exhaust Connection CEC The FlexAIR canopy exhaust connection with integrated alarm provides the cabinet with a safe exhaust connection to building exhaust system...

Page 11: ...BMBL 5th edition published by the U S Department of Health Human Services as HHS Publication No CDC 21 1112 as an advisory document for safe work practices Baker biosafety cabinets are designed for c...

Page 12: ...e device It is recommended that the GFCI device be tested periodically to assure safe operation Ultraviolet UV Light Optional WARNING UV light is hazardous DO NOT defeat the interlock Eyes and skin sh...

Page 13: ...le Proximity Sensor Fault Alarm The Double Proximity Sensor Fault alarm occurs when both Sash Position proximity sensors are activated simultaneously This is a condition that should not be able to occ...

Page 14: ...about an additional 3 seconds you will hear a longer beep This indicates that you have turned the timer function ON are in the programming mode for the device and have programmed it to turn OFF in 1...

Page 15: ...o they cannot operate simultaneously see also step 2 5 Wipe down the interior area of the cabinet with surface disinfectant NOTE Some disinfectants such as bleach or iodine may corrode or stain the st...

Page 16: ...s in and out of the cabinet through the air barrier Room airflow can significantly affect cabinet operation Opening and closing doors in the laboratory can cause air disturbances which might interfere...

Page 17: ...ol or steel pads when cleaning stainless steel Dirt deposits on stainless steel dust dirt and finger marks can usually be removed using warm water with or without detergent If this does not remove the...

Page 18: ...riate agent prior to maintenance service or repairs in any contaminated area of the cabinet The National Institute of Health National Cancer Institute and the Centers for Disease Control have all reco...

Page 19: ...shorten the filter s life The High Efficiency Particulate Air HEPA filter is one of the essential components of a clean air cabinet It is the shield which stands between the product and the environmen...

Page 20: ...he Baker Company Inc will F O B Sanford Maine furnish without charge repairs to or replacement parts or equipment which proved defective in material or workmanship No claim may be made for any inciden...

Reviews: