background image

 
 

389D000 Rev B 

9

 

 

 

Alarm Conditions 

Any alarm condition indicates a potential hazardous condition that needs to be addressed immediately. 
The alarms below are listed in order of precedence.

 

Sash Alarm 

There are two safe sash levels defined for this cabinet, normal operating position and fully closed position. For 
normal operation, the viewscreen sash must be placed at the safe design opening indicated by marker arrows on 
both the right and left sash guide. The sash alarm occurs when the view screen is not at a safe operating 
position, and alerts the user with a visual and audible alarm once per second. There is a 3 second delay before 
the alarm activates, to allow the operator time to move between safe operating positions. Pressing the Alarm 
Reset pushbutton will mute the audible alarm for 5 minutes. The visual alarm will continue to flash until the 
viewscreen is placed at safe level. If the unsafe sash position is not corrected, the audible alarm will return. 

Power/Processor Fault Alarm 

The Power/Processor Fault Alarm occurs when the system experiences a power outage or the 
microprocessor/controller has a fault. The indication of this fault is a visual and audible alarm of three, one 
second alarms followed by a two second delay. This cycle is repeated until the alarm condition is cleared by 
pressing the Alarm Reset pushbutton. 

Double Proximity Sensor Fault Alarm 

The Double Proximity Sensor Fault alarm occurs when both Sash Position proximity sensors are activated 
simultaneously. This is a condition that should not be able to occur in normal operation. A visual and audible 
alarm four times per second indicates this fault condition. This alarm can be muted for five minute periods by 
pressing the Alarm Reset pushbutton. 

Low-flow Exhaust Alarm 

(with FlexAIR canopy exhaust connection only)

 

The

 

Low-flow alarm, (optional components required) occurs when there is a loss of capture air flowing into the 

canopy air intake openings due to low exhaust air flow. A visual and audible alarm, twice per second, will alert 
users of this undesirable operating condition. The Low-flow alarm can be muted, however a visual alarm will 
remain until the problem is resolved. If the FlexAIR exhaust system has not been corrected, the audible alarm 
will return.  

Alarm Summary Table 

Alarm Type 

Acknowledgement 

Comments 

Action 

Sash Alarm 

Once per second after 

three second delay 

Takes precedence over other 

alarms; audible alarm can be 

muted for 5 minutes 

Return sash to correct 

position 

Power/Processor Fault 

Alarm 

Three one second 

alarms followed by two 

second delay 

Cycle will repeat until cleared 

by pushing alarm reset button 

Check power distribution 

to cabinet 

Double Proximity 

Sensor Fault Alarm 

Four times per second 

Can be muted for  5 minutes 

Return sash to one of the 

two correct positions 

Low-flow Exhaust 

Alarm 

Twice per second 

Audible alarm can be muted 

Check building exhaust 

system 

 

 

Summary of Contents for SterilGARD e3 Series

Page 1: ...B i OPERATOR S MANUAL Class II Type A2 Biological Safety Cabinet MODELS SG404 SG504 SG604 SG404 INT SG504 INT SG604 INT SG404 J SG504 J SG604 J Revision Date Description B 11 12 14 Changed Warranty St...

Page 2: ...ring a contamination free atmosphere In addition to the high quality you expect from all Baker equipment this model has been designed to provide the lab user with the most comfortable work environment...

Page 3: ...ea Lighting 4 Electronic Ballast 4 Ground Fault Circuit Interrupter 115V only 4 Laboratory services 4 Alarms 4 Timers 4 Padded Bypass Armrest 5 UniPressure Plenum 5 Motor Blower Assembly 5 Cable Port...

Page 4: ...increment programming 10 1 hour increment programming 10 Cable Ports 10 Cable Ports continued 11 Start up Procedure 11 Working in the Cabinet 12 Using Ancillary Equipment 13 Reacting to Spills 13 Clea...

Page 5: ...supply air passes through a diffuser but some of the air is directed down the back side of the view screen creating a high velocity air curtain at the front access opening The HEPA filtered supply air...

Page 6: ...ILTER POSITIVE PRESSURE PLENUM SUPPLY HEPA FILTER REMOVABLE WORK SURFACE AND SUPPORTS FLUORESCENT LAMP LIGHT CANOPY SLIDING VIEWSCREEN 10 DEGREE SLOPE SUPPLY DIFFUSER FILTER PROTECTOR AUXILIARY OUTLET...

Page 7: ...cabinet has a manual internal balancing damper that compensates for down flow and exhaust intake imbalances WARNING This damper must only be adjusted by a qualified technician who has had proper train...

Page 8: ...The SterilGARD e3 standard view screen access opening is 8 203 mm Other optional factory configurations available with NSF listing are 10 254 mm and 12 305 mm access openings Work Area Lighting The w...

Page 9: ...the recorded value in the most recent certification report However if drastic change is noted this should be cause for investigating the reason This device is not intended to be used for air flow set...

Page 10: ...rnia OSHPD pre approval FlexAIR Canopy Exhaust Connection CEC The FlexAIR canopy exhaust connection with integrated alarm provides the cabinet with a safe exhaust connection to building exhaust system...

Page 11: ...BMBL 5th edition published by the U S Department of Health Human Services as HHS Publication No CDC 21 1112 as an advisory document for safe work practices Baker biosafety cabinets are designed for c...

Page 12: ...e device It is recommended that the GFCI device be tested periodically to assure safe operation Ultraviolet UV Light Optional WARNING UV light is hazardous DO NOT defeat the interlock Eyes and skin sh...

Page 13: ...le Proximity Sensor Fault Alarm The Double Proximity Sensor Fault alarm occurs when both Sash Position proximity sensors are activated simultaneously This is a condition that should not be able to occ...

Page 14: ...about an additional 3 seconds you will hear a longer beep This indicates that you have turned the timer function ON are in the programming mode for the device and have programmed it to turn OFF in 1...

Page 15: ...o they cannot operate simultaneously see also step 2 5 Wipe down the interior area of the cabinet with surface disinfectant NOTE Some disinfectants such as bleach or iodine may corrode or stain the st...

Page 16: ...s in and out of the cabinet through the air barrier Room airflow can significantly affect cabinet operation Opening and closing doors in the laboratory can cause air disturbances which might interfere...

Page 17: ...ol or steel pads when cleaning stainless steel Dirt deposits on stainless steel dust dirt and finger marks can usually be removed using warm water with or without detergent If this does not remove the...

Page 18: ...riate agent prior to maintenance service or repairs in any contaminated area of the cabinet The National Institute of Health National Cancer Institute and the Centers for Disease Control have all reco...

Page 19: ...shorten the filter s life The High Efficiency Particulate Air HEPA filter is one of the essential components of a clean air cabinet It is the shield which stands between the product and the environmen...

Page 20: ...he Baker Company Inc will F O B Sanford Maine furnish without charge repairs to or replacement parts or equipment which proved defective in material or workmanship No claim may be made for any inciden...

Reviews: