background image

Copyright 1998 Baker Hughes Company.

English–DPI610A Instruction Manual | 9

If necessary, SELECT AERO UNITS:

2.2.1  Method

1. Connect the instrument to the static system and, if required, apply a pressure correction 

value of the pressure of the day (POD) as detailed on page

 

6.

2. Close the release valve and manual let-down valve, using the hand-pump, generate the 

required altitude.

3. Set WAIT and DURATION of the leak test times:

4. When ready, press the RUN key (F2):

Note: 

Initially, the WAIT period counts-down, followed by the TIMING period. The START 

ALT is recorded as soon as the timing period starts. At the end of the test the display shows 

the complete details of the test.

5. Record the displayed operating data.
6. To perform another leak test, use the hand-pump to generate a new system altitude and 

press RUN (F2) again.

7. On completion of the test, allow the altitude to decrease until it is below 1500

 

ft.

8. Slowly open the manual let-down valve to return the system to ground level.
9. To exit or cancel leak test mode, press TASK.

2.3  Airspeed Switch Tests

Select TASK, AERO:

SELECT AERO

UNITS

ft/minute

m/minute
ft/second
m/second
hm/minute

SELECT UNITS

INPUT

RUN

CHANGE 

VALUE

ALTITUDE

WAIT 

60

 secs

DURATION 60  secs

START ALT. 

 

ft

STOP ALT. 

 

ft

ALT. CHANGE 

 

ft

LEAK RATE 

 

ft/min

ft

ft

ft

RUN

CHANGE 

VALUE

ALTITUDE

WAIT 

60

 secs

DURATION 60  secs

START ALT. 

 

ft

STOP ALT. 

 

ft

ALT. CHANGE 

 

ft

LEAK RATE 

 

ft/min

SET WAIT TIME

SET DURATION

RUN

CHANGE 

VALUE

ALTITUDE

WAIT 60 

secs

DURATION 

60

 secs

START ALT. 

 

ft

STOP ALT. 

 

ft

ALT. CHANGE 

 

ft

LEAK RATE 

 

ft/min

ft

ft

TIMING

ft

WAITING

ALTITUDE

ALTITUDE

POD = 1002.85

POD = 1002.85

STOP

STOP

RUN

CHANGE

VALUE

ALTITUDE

WAIT 60 

secs

DURATION 

60

 secs

START ALT. 

13354 

ft

STOP ALT. 

13322 

ft

ALT. CHANGE 

-32 

ft

LEAK RATE 

-32 

ft/min

Summary of Contents for Druck DPI 610A

Page 1: ...Druck com DPI 610A Aeronautical Portable Pressure Calibrator Instruction Manual...

Page 2: ......

Page 3: ...acturer s service departments Druck com For technical advice contact the manufacturer Symbols WARNING Before operating this instrument read the safety instructions Do not apply pressures greater than...

Page 4: ...18 ft 30 000 ft 36 ft Precision includes linearity hysteresis and repeatability 12 months measurement stability Temperature Effects 0 004 of reading C averaged over 10 to 40 C w r t 20 C 0 002 of read...

Page 5: ...g absolute version of the DPI 610 provides manual generation of pressure and vacuum for testing of aircraft Pitot and Static systems It has a special pneumatic manifold and output port assembly the re...

Page 6: ...ator Controls 1 Output port AN4 to pitot or static system 2 Manual let down valve 3 Electrical switch inputs 4 Release valve releases system pressure vacuum at a controlled rate 5 Vent port 6 Selects...

Page 7: ...imum measurement ratings of the standard instrument that should not be exceeded Pressure over range 120 of FS is indicated by a flashing pressure display Voltage or current over range 110 of max cause...

Page 8: ...n POD by selecting POD ADJUST as follows Apply QFE pressure reading as POD correction as follows WARNING Observe safety precautions stated in local orders and the aircraft or equipment servicing proce...

Page 9: ...RUN key F1 and operate the hand pump continue until the switch operates mbar POD Off POD Correction mbar POD Correction POD Off WARNING VENT SYSTEM BEFORE SETTING POD WARNING VENT SYSTEM BEFORE SETTIN...

Page 10: ...m is vented to atmosphere before starting the test 2 Select negative pressure and close the release valve 3 Close the manual let down valve 4 Press the RUN key F1 and operate the hand pump continue un...

Page 11: ...a new system altitude and press RUN F2 again 7 On completion of the test allow the altitude to decrease until it is below 1500 ft 8 Slowly open the manual let down valve to return the system to ground...

Page 12: ...be potential free Figure 4 Airspeed Switch Test Connections 2 Ensure that positive pressure is selected and close the release valve 3 Close the manual let down valve 4 Zero the airspeed reading by pre...

Page 13: ...ng the test 2 Ensure that positive pressure is selected and close the release valve 3 Close the manual let down valve 4 Zero the airspeed reading by pressing the ZERO key 5 Press the RUN key F1 and op...

Page 14: ...layed operating data 6 Allow the airspeed to decrease until it is below 80 kts 7 Slowly open the manual let down valve to return the system to zero ground level SELECT TASK SELECT TASK ALTITUDE SWITCH...

Page 15: ...1998 Baker Hughes Company English DPI610A Instruction Manual 13 8 To perform another leak test use the hand pump to generate a new airspeed and press RUN F2 again 9 To exit or cancel leak test mode p...

Page 16: ...Copyright 1998 Baker Hughes Company 14 DPI610A Instruction Manual English...

Page 17: ......

Page 18: ......

Page 19: ......

Page 20: ...kerhughes com Bangalore 91 9986024426 aneesh madhav bakerhughes com Milan 39 02 36 04 28 42 csd italia bakerhughes com Japan Netherlands Russia Tokyo 81 3 6890 4538 gesitj bakerhughes com Hoevelaken 3...

Reviews: